Preparation for operation
21
Leica M525 OH4 / Ref. 10 714 367 / Version -
Checklist: Before the operation
Cleaning the optical accessories
➩
Check the tubes, eyepieces and the documentation
accessories (if used) for cleanliness.
➩
Remove dust and dirt (see page 68).
Installing accessories
➩
Lock the Leica M525 OH4 in place and install all accessories
on the microscope so it is ready for use (see page 22).
➩
Position the CAN handles as required.
➩
Connect a mouth switch and/or footswitch (if used).
➩
Check the camera image on the monitor and realign if
necessary.
Checking tube settings
➩
Check the tube and eyepiece setting for the selected user
(see page 24).
➩
Treat the eyepieces with an antifogging compound if
necessary.
Balancing
➩
Balancing the Leica M525 OH4 (see page 27).
➩
Press the "All Brakes" button on the CAN handle and check
the balancing.
Function check
Warning 1
Danger of fatal electrical shock.
➩
The Leica M525 OH4 surgical microscope may be
connected to a grounded socket only.
➩
Connect the power cable.
➩
Switch on the microscope.
➩
Switch on the illumination at the control unit.
Leave the illumination on for at least 5 minutes, as
otherwise luminosity will decrease rapidly.
➩
Replace defective bulbs before the operation begins.
➩
Test all functions on the CAN handles and on the footswitch.
➩
Check the user settings on the control unit for the selected
user (see page 45).
Positioning at the OP table
➩
Position the surgical microscope on the OP table as required,
and lock the footbrake (see page 35).
Sterility
➩
Fit sterile components and sterile drape if used (see page 36).
Warning 2
Danger of fatal electric shock
➩
Operate the system only with all equipment in its
proper position (all covers fitted, doors closed).
Summary of Contents for OH4
Page 2: ......
Page 12: ...Releases Leica OH4 Service manual Releases 04 2007 Edition 1...
Page 14: ...Technical description Service manual Leica OH4 09 2006 3243 GOR 2 II Doc Code 10 665 742...
Page 26: ...Technical Description Service manual Leica OH4 9 2006 3243 GOR 1 12 Doc Code 10 665 744...
Page 28: ...Wiring Service manual Leica M525 OH4 04 2007 Gahe 02 II Doc Code 10665744...
Page 29: ...Service manual Leica M525 OH4 Wiring Doc Code 10665744 2 1 04 2007 Gahe...
Page 30: ...Wiring Service manual Leica M525 OH4 04 2007 Gahe 2 2 Doc Code 10665744...
Page 31: ...Service manual Leica M525 OH4 Wiring Doc Code 10665744 2 3 04 2007 Gahe...
Page 32: ...Wiring Service manual Leica M525 OH4 04 2007 Gahe 2 4 Doc Code 10665744...
Page 33: ...Service manual Leica M525 OH4 Wiring Doc Code 10665744 2 5 04 2007 Gahe...
Page 34: ...Wiring Service manual Leica M525 OH4 04 2007 Gahe 2 6 Doc Code 10665744...
Page 36: ......
Page 38: ...Service Programs Service manual Leica OH4 11 2006 3243 GOR Doc Code 10 665 744 3 II...
Page 96: ...Maintenance service Service manual Leica OH4 stand 04 2007 3243 GOR 6 2 Doc Code 10 665 744...
Page 98: ...Inspection Service manual Leica OH4 stand 04 2007 3243 GOR 7 II Doc Code 10 665 744...
Page 102: ...Inspection Service manual Leica OH4 stand 04 2007 3243 GOR 7 4 Doc Code 10 665 744...
Page 141: ...Leica M525 OH4 User manual 10 714 367 version...
Page 223: ...Technical data 81 Leica M525 OH4 Ref 10 714 367 Version 1200 max 1925 720 x 720 2685 1945...
Page 224: ...Technical data 82 Leica M525 OH4 Ref 10 714 367 Version 1180 1785 430...
Page 225: ...Technical data 83 Leica M525 OH4 Ref 10 714 367 Version...
Page 226: ...Notes 84 Leica M525 OH4 Ref 10 714 367 Version...
Page 227: ...Notes 85 Leica M525 OH4 Ref 10 714 367 Version...
Page 229: ...Leica M525 OH4 Assembly instructions 10 714 368 Version A...
Page 230: ...Leica M525 OH4 Ref 10 714 368 Version A...