Use
42
Leica M525 OH4 / Ref. 10 714 367 / Version -
Setting the illumination field diameter
Caution 7
Heating of tissue!
If the illumination field diameter is larger than the
visual field and the light intensity is set too high, uncon-
trolled heating of tissue can occur outside the visible
range of the microscope.
➩
Do not set the light intensity too high.
The light field diameter is automatically adapted to the size of
the visual field at the Leica M525 optical carrier. The
illumination field diameter can also be adjusted manually with
rotary button (2). Automatic adjustment can be deactivated with
the rotary button and reactivated with the Reset button (1).
If the illumination field diameter is blocked at a high
light intensity in a high zoom setting, and cannot be
adjusted automatically or manually, then the light
intensity must be reduced in order to protect the tissue.
If the illumination field diameter remains blocked in a
low zoom setting, and cannot be adjusted
automatically or manually, then an OP lamp can be
used for better illumination of a large visual field (low
zoom setting).
2
1
Adjusting the magnification (zoom)
You can set the magnification with a footswitch/handswitch or
with the "Magnification" adjustment bar in the "Main" menu of
the control unit.
Clicking the "+" or "-" key changes the magnification
value in increments of one. Holding down the mouse
button with your finger changes the value in
increments of five.
You can adjust the zoom motor speed in the "Speed"
menu.
These values can be saved individually for each user
(see page 46).
Warning 8
Danger to the patient due to failure of the zoom motor!
➩
In case of failure of the zoom motor, the zoom can be
adjusted manually.
Manually adjusting the magnification (zoom)
Caution 8
Destruction of the zoom motor!
➩
Use the manual adjustment of the zoom motor only if
the zoom motor is defective.
In case of failure of the zoom motor, the zoom can be adjusted
manually using rotary button (3).
➩
Hold down rotary button (3).
➩
Set the desired magnification by turning the knob.
3
Summary of Contents for OH4
Page 2: ......
Page 12: ...Releases Leica OH4 Service manual Releases 04 2007 Edition 1...
Page 14: ...Technical description Service manual Leica OH4 09 2006 3243 GOR 2 II Doc Code 10 665 742...
Page 26: ...Technical Description Service manual Leica OH4 9 2006 3243 GOR 1 12 Doc Code 10 665 744...
Page 28: ...Wiring Service manual Leica M525 OH4 04 2007 Gahe 02 II Doc Code 10665744...
Page 29: ...Service manual Leica M525 OH4 Wiring Doc Code 10665744 2 1 04 2007 Gahe...
Page 30: ...Wiring Service manual Leica M525 OH4 04 2007 Gahe 2 2 Doc Code 10665744...
Page 31: ...Service manual Leica M525 OH4 Wiring Doc Code 10665744 2 3 04 2007 Gahe...
Page 32: ...Wiring Service manual Leica M525 OH4 04 2007 Gahe 2 4 Doc Code 10665744...
Page 33: ...Service manual Leica M525 OH4 Wiring Doc Code 10665744 2 5 04 2007 Gahe...
Page 34: ...Wiring Service manual Leica M525 OH4 04 2007 Gahe 2 6 Doc Code 10665744...
Page 36: ......
Page 38: ...Service Programs Service manual Leica OH4 11 2006 3243 GOR Doc Code 10 665 744 3 II...
Page 96: ...Maintenance service Service manual Leica OH4 stand 04 2007 3243 GOR 6 2 Doc Code 10 665 744...
Page 98: ...Inspection Service manual Leica OH4 stand 04 2007 3243 GOR 7 II Doc Code 10 665 744...
Page 102: ...Inspection Service manual Leica OH4 stand 04 2007 3243 GOR 7 4 Doc Code 10 665 744...
Page 141: ...Leica M525 OH4 User manual 10 714 367 version...
Page 223: ...Technical data 81 Leica M525 OH4 Ref 10 714 367 Version 1200 max 1925 720 x 720 2685 1945...
Page 224: ...Technical data 82 Leica M525 OH4 Ref 10 714 367 Version 1180 1785 430...
Page 225: ...Technical data 83 Leica M525 OH4 Ref 10 714 367 Version...
Page 226: ...Notes 84 Leica M525 OH4 Ref 10 714 367 Version...
Page 227: ...Notes 85 Leica M525 OH4 Ref 10 714 367 Version...
Page 229: ...Leica M525 OH4 Assembly instructions 10 714 368 Version A...
Page 230: ...Leica M525 OH4 Ref 10 714 368 Version A...