70
Leica M525 OH4 / Ref. 10 714 367 / Version -
Notes on reprocessing of resterilizable
products
Products
Reusable products supplied by Leica Microsystems (Schweiz)
AG such as rotary knobs, objective protective glasses and
capping pieces.
Limitation of reprocessing:
For the medical devices used on patients suffering from
Creutzfeldt Jacob Disease (CJD) or suspected of having CJD or
variant CJD, the local statutory requirements have to be met.
Normally resterilizable products used on this group of patients
are to be eliminated without risk by incineration.
Occupational safety and health protection
Particular attention must be paid to the occupational safety and
health protection of the persons responsible for preparing cont-
aminated products. Current regulations of hospital hygiene and
prevention of infection must be observed in the preparation,
cleaning and disinfection of the products.
Limitation of reprocessing
Frequent reprocessing has little effects on these products.
The end of the product life cycle is usually determined by wear
and year and damage through use.
Instructions
Workplace
Remove surface contamination with a disposable cloth / paper
cloth.
Storage and transport
No special requirements.
It is recommended to perform the reprocessing of a product
immediately following its use.
Preparation for cleaning
Remove the product from the Leica M525 OH4 surgical
microscope.
Cleaning: manual
Equipment: running water, rinsing agent, spirit, micro-fiber cloth
Procedure:
1. Rinse surface contamination off of the product (temp. <40°C).
Use some rinsing agent depending upon degree of
contamination.
2. If the optics is heavily contaminated, e.g. finger marks, fat
streaks etc., use spirit for cleaning.
3. Dry off products, except for optical components, with a
disposable cloth/paper cloth. Dry off optical surfaces with a
micro-fiber cloth.
Cleaning: automatic
Equipment: Cleaning/disinfecting device
It is not recommended to clean products with optical
components in a cleaning/disinfecting device. In addition,
optical components must not be cleaned in ultrasonic baths in
order to prevent damage.
Disinfection
The alcohol disinfection solution "Mikrozid, Liquid" may be used
in accordance with the instructions on the label.
Please note that optical surfaces must be rinsed thoroughly
with fresh drinking water, followed by fresh demineralized
water, after disinfection. The products must be dried thoroughly
before the subsequent sterilization.
Maintenance
No special requirements.
Control and functional test
Check the snap-on behavior of rotary knobs and CAN handles.
Packaging
Separate: A standard PE bag may be used. The bag must be
large enough for the product so that the closure is not under
tension.
Sterilization
See Table 1
Storage
No special requirements.
Additional information
None
Contact information of manufacturer
Address of local agent
Leica Microsystems (Schweiz) AG verified that the
aforementioned instructions for the preparation of a product
are suitable for its reuse. The processing person is responsible
for reprocessing with the equipment, materials and personnel
and for achieving the desired results in the reprocessing instal-
lation. In general, this requires validations and routine monitor-
ing of the process. Every deviation from the supplied
instructions should also be examined carefully by the
processing person to determine effectiveness and possible
detrimental consequences.
Care and maintenance
Summary of Contents for OH4
Page 2: ......
Page 12: ...Releases Leica OH4 Service manual Releases 04 2007 Edition 1...
Page 14: ...Technical description Service manual Leica OH4 09 2006 3243 GOR 2 II Doc Code 10 665 742...
Page 26: ...Technical Description Service manual Leica OH4 9 2006 3243 GOR 1 12 Doc Code 10 665 744...
Page 28: ...Wiring Service manual Leica M525 OH4 04 2007 Gahe 02 II Doc Code 10665744...
Page 29: ...Service manual Leica M525 OH4 Wiring Doc Code 10665744 2 1 04 2007 Gahe...
Page 30: ...Wiring Service manual Leica M525 OH4 04 2007 Gahe 2 2 Doc Code 10665744...
Page 31: ...Service manual Leica M525 OH4 Wiring Doc Code 10665744 2 3 04 2007 Gahe...
Page 32: ...Wiring Service manual Leica M525 OH4 04 2007 Gahe 2 4 Doc Code 10665744...
Page 33: ...Service manual Leica M525 OH4 Wiring Doc Code 10665744 2 5 04 2007 Gahe...
Page 34: ...Wiring Service manual Leica M525 OH4 04 2007 Gahe 2 6 Doc Code 10665744...
Page 36: ......
Page 38: ...Service Programs Service manual Leica OH4 11 2006 3243 GOR Doc Code 10 665 744 3 II...
Page 96: ...Maintenance service Service manual Leica OH4 stand 04 2007 3243 GOR 6 2 Doc Code 10 665 744...
Page 98: ...Inspection Service manual Leica OH4 stand 04 2007 3243 GOR 7 II Doc Code 10 665 744...
Page 102: ...Inspection Service manual Leica OH4 stand 04 2007 3243 GOR 7 4 Doc Code 10 665 744...
Page 141: ...Leica M525 OH4 User manual 10 714 367 version...
Page 223: ...Technical data 81 Leica M525 OH4 Ref 10 714 367 Version 1200 max 1925 720 x 720 2685 1945...
Page 224: ...Technical data 82 Leica M525 OH4 Ref 10 714 367 Version 1180 1785 430...
Page 225: ...Technical data 83 Leica M525 OH4 Ref 10 714 367 Version...
Page 226: ...Notes 84 Leica M525 OH4 Ref 10 714 367 Version...
Page 227: ...Notes 85 Leica M525 OH4 Ref 10 714 367 Version...
Page 229: ...Leica M525 OH4 Assembly instructions 10 714 368 Version A...
Page 230: ...Leica M525 OH4 Ref 10 714 368 Version A...