Use
57
Leica M525 OH4 / Ref. 10 714 367 / Version -
The Leica M525 OH4 surgical microscope is state-of-the-art
technology. Nevertheless, hazards can arise during operation.
➩
Always follow the instructions in this user manual, and in par-
ticular the safety guidelines.
Intended use
• The Leica M525 OH4 surgical microscope is an optical instru-
ment for improving the visibility of objects through magnifica-
tion and illumination. It can be applied for observation and
documentation and for human and veterinary medical
treatment.
• The Leica M525 OH4 surgical microscope may be used only in
closed rooms and must be placed on a solid floor.
• The Leica M525 OH4 surgical microscope is subject to
special precautionary measures for electromagnetic compat-
ibility. They must be installed and put into operation in accor-
dance with the guidelines, manufacturer's declarations and
recommended safety distances (Tables 201, 202, 204, and 206
according to EN 60601-1-2:2001).
• Portable and mobile as well as stationary RF communications
equipment can have a negative effect on the reliability of the
Leica M525 OH4 surgical microscope.
Warning 12
Not suitable for use in ophthalmology!
Directions for the person responsible for
the instrument
➩
Ensure that the Leica M525 OH4 surgical microscope is used
only by persons qualified to do so.
➩
Ensure that this user manual is always available at the place
where the Leica M525 OH4 surgical microscope is in use.
➩
Carry out regular inspections to make certain that the
authorized users are adhering to safety requirements.
➩
When instructing new users, do so thoroughly and explain
the meanings of the warning signs and messages.
➩
Assign individual responsibilities for starting, operating and
servicing the Leica surgical microscope and monitor the
observance of these responsibilities.
➩
Only use the Leica M525 OH4 surgical microscope if it is free
of defects.
➩
Inform your Leica representative or Leica Microsystems
(Schweiz) AG, BU SOM, 9435 Heerbrugg, Switzerland,
immediately if you detect a product defect that could
potentially cause injury or harm.
➩
If you use accessories made by third-party manufacturers
with the Leica M525 OH4 surgical microscope, be sure that
each such manufacturer confirms the safety-engineering,
harmless usability of the product and observe the product's
user manual.
➩
Modifications to or service on the Leica M525 OH4 surgical
microscope may be carried out only by technicians who are
explicitly authorized by Leica to do so.
➩
Only original Leica replacement parts may be used in servic-
ing the product.
➩
After servicing or making technical modifications, the instru-
ment must be re-set in accordance with our technical specifi-
cations.
➩
Leica accepts no liability if the instrument is modified or
serviced by unauthorized persons, not maintained as
specified (unless maintenance is performed by us) or not
used/handled properly.
➩
Interaction of other devices through the Leica surgical micro-
scope was proven in accordance with EN 60 601-1-2. The
system passed the emission and immunity test. The standard
safety precautions and safety requirements relating to
electromagnetic and other radiation must be observed.
➩
Domestic electrical installation must be conducted in
accordance with the national standard, e.g. use of a residual
current protective device (FI protection) is recommended.
➩
As with every other instrument used in operative
applications, this system is subject to failure. Leica
Microsystem (Schweiz) AG therefore recommends that a
backup system be kept available during the operation.
Summary of Contents for OH4
Page 2: ......
Page 12: ...Releases Leica OH4 Service manual Releases 04 2007 Edition 1...
Page 14: ...Technical description Service manual Leica OH4 09 2006 3243 GOR 2 II Doc Code 10 665 742...
Page 26: ...Technical Description Service manual Leica OH4 9 2006 3243 GOR 1 12 Doc Code 10 665 744...
Page 28: ...Wiring Service manual Leica M525 OH4 04 2007 Gahe 02 II Doc Code 10665744...
Page 29: ...Service manual Leica M525 OH4 Wiring Doc Code 10665744 2 1 04 2007 Gahe...
Page 30: ...Wiring Service manual Leica M525 OH4 04 2007 Gahe 2 2 Doc Code 10665744...
Page 31: ...Service manual Leica M525 OH4 Wiring Doc Code 10665744 2 3 04 2007 Gahe...
Page 32: ...Wiring Service manual Leica M525 OH4 04 2007 Gahe 2 4 Doc Code 10665744...
Page 33: ...Service manual Leica M525 OH4 Wiring Doc Code 10665744 2 5 04 2007 Gahe...
Page 34: ...Wiring Service manual Leica M525 OH4 04 2007 Gahe 2 6 Doc Code 10665744...
Page 36: ......
Page 38: ...Service Programs Service manual Leica OH4 11 2006 3243 GOR Doc Code 10 665 744 3 II...
Page 96: ...Maintenance service Service manual Leica OH4 stand 04 2007 3243 GOR 6 2 Doc Code 10 665 744...
Page 98: ...Inspection Service manual Leica OH4 stand 04 2007 3243 GOR 7 II Doc Code 10 665 744...
Page 102: ...Inspection Service manual Leica OH4 stand 04 2007 3243 GOR 7 4 Doc Code 10 665 744...
Page 141: ...Leica M525 OH4 User manual 10 714 367 version...
Page 223: ...Technical data 81 Leica M525 OH4 Ref 10 714 367 Version 1200 max 1925 720 x 720 2685 1945...
Page 224: ...Technical data 82 Leica M525 OH4 Ref 10 714 367 Version 1180 1785 430...
Page 225: ...Technical data 83 Leica M525 OH4 Ref 10 714 367 Version...
Page 226: ...Notes 84 Leica M525 OH4 Ref 10 714 367 Version...
Page 227: ...Notes 85 Leica M525 OH4 Ref 10 714 367 Version...
Page 229: ...Leica M525 OH4 Assembly instructions 10 714 368 Version A...
Page 230: ...Leica M525 OH4 Ref 10 714 368 Version A...