7
5.4
MOVE DOWN
Lowers the stereomicroscope (move command)
Command code:
70 011 <CR>
Parameter:
none
Return:
none
Comment:
Lowers the stereomicroscope at the set speed down to the lower limit switch or lower limit
(see also MOVE UP).
5.5
MOVE STOP
Interrupts movement, stops the stereomicroscope.
Command code:
70 012
Parameter:
none
Return:
none
Comment:
Stops the movement of the stereomicroscope, the stereomicroscope is braked with the
maximum deceleration.
5.6
SET REFERENCE (set zero reference point)
Sets the zero reference point at the momentary position
Command code:
70 013
Parameter:
none
Return:
none
Comment:
The zero reference point is set at the momentary position. If any limits and positions are
already detected or have been set they will be adjusted for the difference compared with
the old zero point.
5.7.
MOVE ABSOLUT
Moves to the absolute position (move command)
Command code:
70 030
Parameter:
10 bytes
Position to be adopted
Range:
±
2.147.483.647 (long int - 32 bit)