²
0HDVXULQJ6NLOOV
Now in Sequence Mode proper, the screen displays:
Note: The number of trains
acquired is now four, with the
same sampling rate as for
two. The sole difference is
that now the duration between
two train pulses is half what it
was. In the trace label, at left-
of-screen, the symbol “2x”
indicates two segments
acquired in Sequence Mode.
The TIMEBASE information
shows that two segments of
50 000 points each
(“2
F
50 000”) have been
acquired.
7LSV
: The bottom menu shows the maximum number of
points per segment — in this instance, 50 000. The sampling
rate or the total number of acquired points can be adjusted.
The timebase setting in Sequence Mode is used for
determining the acquisition duration of each segment —
10 x TIME/DIV. This setting, as well as the number of
segments desired, the maximum segment length, and the
total memory available, are used for arriving at the actual
number of samples per segment, and the time per point.