Table 37: Cooling circulation thermostats and heating circulation thermostats
Unit
RP 240 E
RP 245 E
RP 250 E
RP 290 E
P 2 E
*ACC area/working tem-
perature range
°C
-40 – 200* -45 – 200*
-50 – 200* -90 – 200*
80 – 250
Operating temperature
range (heating device with
external cooling)
°C
---
---
---
---
-30 – 250
Device dimensions (W x D)
mm
300 x 430
300 x 430
300 x 430
390 x 600
250 x 365
Device height (H), including
Command Touch unit
mm
675
675
675
685
425
Filling volume
‑
maximum
L
4.4
4.4
4.4
4.4
4.4
‑
minimum
L
2.4
2.4
2.4
2.4
2.4
Pump data
Maximum discharge pres-
sure
bar
0.7
0.7
0.7
0.7
0.7
Maximum pump suction
bar
0.4
0.4
0.4
0.4
0.4
Maximum flow rate pressure
l/min
22
22
22
22
22
Maximum flow rate suction
l/min
20
20
20
20
20
Pump connecting thread
mm
M16 x 1
M16 x 1
M16 x 1
M16 x 1
M16 x 1
Connecting thread on
cooling coil (thermal devices
only)
Inches
---
---
---
---
Thread G3/8"
external &
G1/4" internal
Noise level (1 m)
dB(A)
54
54
57
56
47
Weight
kg
46
46
47
79
16
Clearance
‑
Front
mm
200
200
200
200
200
‑
Back
mm
200
200
200
200
200
‑
Right
mm
200
200
200
200
200
‑
Left
mm
200
200
200
200
200
*ACC area (Active Cooling Control) according to DIN 12876 is
the working temperature range during operation with an active
refrigerating machine.
11.2
Cooling capacity and cooling water
The devices are operated with partially halogenated and/or natural
refrigerant, depending on the device model. The cooling output
values measured for partially halogenated refrigerant and natural
refrigerant are identical. The designation and refrigerant charge are
specified on the device type plate.
V6
PRO bath thermostats and circulation thermostats
133 / 156
Summary of Contents for P 10
Page 146: ...V6 PRO bath thermostats and circulation thermostats 146 156 ...
Page 147: ...V6 PRO bath thermostats and circulation thermostats 147 156 ...
Page 148: ...V6 PRO bath thermostats and circulation thermostats 148 156 ...
Page 157: ......
Page 158: ......
Page 159: ......