For internal bath applications, the reference thermometer must be hung
into the bath according to the specifications on the calibration certifi-
cate.
For external applications, the reference thermometer must be installed
in the inlet of the device according to the specifications on the calibra-
tion certificate.
To measure the temperature, wait until the system is in a steady state.
1.
Open the Home window.
2.
Touch the tile
Settings
Calibration
Internal Pt1000
or
External
Pt100
Offset
.
An input window opens.
3.
Enter the temperature value read off the reference thermometer into
the Command Touch unit.
4.
Press [OK] to confirm the new value.
The new value has been accepted.
For internal bath applications, the reference thermometer must be hung
into the bath according to the specifications on the calibration certifi-
cate.
For external applications, the reference thermometer must be installed
in the inlet of the device according to the specifications on the calibra-
tion certificate.
The upper and lower temperature value must be at least 40 K apart.
To measure the temperature, wait until the system is in a steady state.
1.
Set a low T
set
set point on the device.
2.
Wait until the set point and the temperature of the heat transfer liquid
have equaled out.
3.
Open the Home window.
4.
Touch the tile
Settings
Calibration
Internal Pt1000
or
External
Pt100
2-point lower
.
An input window opens.
5.
Enter the temperature value read off the reference thermometer into
the Command Touch unit.
6.
Press [OK] to confirm the new value.
The lower value has been accepted.
7.
Set a high T
set
set point on the device.
8.
Wait until the set point and the temperature of the heat transfer liquid
have equaled out.
9.
Touch the [2-point upper] button in the
Calibration
menu.
10. Enter the temperature value read off the reference thermometer into
the Command Touch unit.
11.
Press [OK] to confirm the new value.
The upper value has been accepted. 2-point calibration has been
completed.
Offset
2-point calibration
V6
PRO bath thermostats and circulation thermostats
110 / 156
Summary of Contents for P 10
Page 146: ...V6 PRO bath thermostats and circulation thermostats 146 156 ...
Page 147: ...V6 PRO bath thermostats and circulation thermostats 147 156 ...
Page 148: ...V6 PRO bath thermostats and circulation thermostats 148 156 ...
Page 157: ......
Page 158: ......
Page 159: ......