Each program starts with the segment
Start
, which determines the tem-
perature at which segment
1
continues the program. The temperature of
the
Start
segment is reached as quickly as possible. It is not possible to
specify a time limit for the
Start
segment. On thermostats without cooling,
the selected start temperature must be higher than the current bath tem-
perature at the time the program starts. Without the
Start
segment, seg-
ment
1
would be different depending on the bath temperature at program
start.
Table 31: Program example, before (
_____
)
Seg.
Tend°C
Time h:m
D
T°C
Control
Out1
Out2
Out3
Start
30.00
----
-----
1
Internal
-----
-----
-----
1
30.00
0:20
0.10
2
Internal
-----
-----
-----
2
50.00
0:20
-----
3
Internal
-----
-----
-----
3
70.00
0:40
-----
4
Internal
-----
-----
-----
4
70.00
0:10
0.10
2
Internal
-----
-----
-----
5
60.00
0:30
-----
2
Internal
-----
-----
-----
6
40.00
0:00
-----
2
Internal
-----
-----
-----
In the edited table, a new segment with the number 3 has been entered.
The time and pump level for segment number 4 have been modified. The
tolerance and pump level for segment number 5 have been adapted.
Table 32: Program example, after (- - - - dashed line, edited)
Seg.
Tend°C
Time h:m
D
T°C
Control
Out1
Out2
Out3
Start
30.00
-----
-----
1
Internal
-----
-----
-----
1
30.00
0:20
0.10
2
Internal
-----
-----
-----
2
50.00
0:20
-----
3
Internal
-----
-----
-----
3
50.00
0:20
0.10
4
Internal
-----
-----
-----
4
70.00
0:20
-----
3
Internal
-----
-----
-----
5
70.00
0:10
0.80
1
Internal
-----
-----
-----
6
60.00
0:30
-----
2
Internal
-----
-----
-----
7
40.00
0:00
-----
2
Internal
-----
-----
-----
Note the following and see Fig. 56:
The Tolerance field ensures strict compliance with the residence time at
a specific temperature, for example.
The subsequent segment is only processed when the actual temperature
reaches the tolerance range (1) so the ramp in the second segment is
delayed and only starts at 2, for example.
Start segment
Tolerance
V6
PRO bath thermostats and circulation thermostats
94 / 156