ECO GOLD
32
Before putting the device into operation
25/08/2011/ YACE0088
a) Approved heat transfer liquids
LAUDA
designation
Operating
temperature
range
Chemical
designation
Viscos-
ity
(kin)
Viscosity
(kin) at
temperature
Fire
point
Container size
Catalogue number
°C
mm²/s at
20 °C
mm²/s
°C
5 L
10 L
20 L
Kryo 51
-50…120
Silicone oil
5
34 at -50 °C
> 160
LZB 121 LZB 221 LZB 321
Kryo 30
-30…90
Monoethylene
glycol /
water
4
50 at -25 °C
--
LZB 109 LZB 209 LZB 309
Kryo 20
-20…180
Silicone oil
11
28 at -20 °C
> 230
LZB 116 LZB 216 LZB 316
Therm 180
0…180
Silicone oil
23
36 at 0 °C
> 288
LZB 115 LTB 214 LZB 314
Aqua 90
5...90
Decalcified water
1 -- --
LZB 120 LZB 220 LZB 320
Ultra 350
30...200
Synth. heat carrier
47
28 at 30 °C
7
240
LZB 107 LZB 207 LZB 307
Therm 240
50…240
Silicone oil
125
45 at 50 °C
7
378
LZB 122 LZB 222 LZB 322
Therm 200
60...200
Silicone oil
54
28 at 60 °C
7
362
LZB 117 LZB 217 LZB 317
At higher temperatures vaporization losses occur. In this case use a bath cover (
Ì
9). Use distilled water or pure
demineralized water only after adding 0.1 g of soda (Na
2
CO
3
sodium carbonate) per liter of water. Otherwise there
is the risk of corrosion!
The proportion of water reduces with longer working at high temperatures and the mixture becomes flammable
(flash point 128 °C). Check the mixing ratio using a hydrometer.
Do not use in conjunction with EPDM hose.
Never use silicone oil with silicone hoses.
EPDM hose is not suitable for Ultra 350 nor for mineral oils.
−
When choosing the heat transfer liquid, it must be noted that at the lower limit of the operating tempera-
ture range impairment of the heat transfer properties is to be expected due to the increasing viscosity.
Therefore, only use the full operating temperature range where necessary.
−
The working ranges of the heat carrier liquids and hoses are general figures which can be tightened due
to the operating temperature range of the devices.
−
Never use contaminated heat transfer liquids. Contamination of the pump chamber may lead to the pump
jamming and the device then switching off.
−
Pay attention to the safety data sheet for the heat transfer liquid.
−
Follow the regulations for disposal of the used heat transfer liquid.
If required, you can request the safety data sheets at any time.
(
Ì