ECO GOLD
82
Appendix with settings
25/08/2011/
YACE0088
A.2
Setting the volume of the acoustic signals
The ECO GOLD thermostats sound alarms and faults as a two-tone acoustic signal. Warnings a signaled
as a continuous tone,
−
Access to the main menu level is obtained by pressing the
enter key
.
−
Selection and confirmation of
Æ
Setup
Æ
Basic setup
Æ
Sounds
.
−
Choose
Alarm
,
Warn
or
Error
.
The adjacent menu window appears.
−
The volume is selected with or . The selected level is
immediately active without confirmation. In this example the
volume is
medium
.
−
By pressing (ESC), or
you are returned to the menu level without any change.
A.3
Setting the chiller
The chiller of the cooling thermostats is operated in the "automatic" operating mode as standard. Here,
the cooling unit switches on or off automatically depending on the temperature and operating status.
However, you can also switch the cooling unit on or off manually.
−
Access to the main menu level is obtained by pressing the
enter key
.
−
Selection and confirmation of
Æ
Setup
Æ
Cooling
.
The adjacent menu window appears.
(As well as
Cooling
the current operating mode is stated. This
is "off", "on" or "automatic".)
−
Select
Cooling
with
.
−
With or and
you select and confirm the operating
status "off", "on" or "automatic".
−
In the menu the set operating status is displayed by
a tick
.
−
By pressing or (ESC) you are returned to the menu level without any change.
Note:
When the cooling unit is switched off, it can take up to two minutes before it switches on again.
ESC
סּ
OK
Standby
Warn
loud
medium
low
off
ESC
סּ
OK
Standby
Setup
Pump
level
►
Control
►
Cooling autom.
►
Limits
►
Basic
setup
►
Calibration
►
ESC
סּ
OK
Standby
Cooler
off
on
.
automatic