ECO GOLD
30
Before putting the device into operation
25/08/2011/ YACE0088
1
2
Draining and changing the heat transfer liquid
−
Switch off the thermostat and withdraw the mains plug.
−
Allow the device and heat transfer liquid to cool down to or
warm up to room temperature.
−
Push a hose onto the bath drain point.
−
Drain the heat transfer liquid via the drain tap at the back
of the device.
Completely drain the bath, external consumers, accessories and hose connections and flush or clean
them (e.g. with new heat transfer liquid).
Caution !
Contact with hot / cold heat transfer liquid
Scalds, frostbite
•
Bring heat transfer liquids to room temperature be-
fore draining.
•
Make sure that the drain tap is closed after draining.
Caution !
Delay in boiling and thermal decomposition due to liquid
residues
Burns, scalds, development of harmful vapors
•
Remove all old heat transfer liquid completely from
the bath, external consumers, accessories and hos-
es. Flush and clean them with new heat transfer liq-
uid.
1
Drain tap, cooling thermostats
2
Drain tap, heating thermostats