ECO GOLD
114
Appendix with settings
25/08/2011/
YACE0088
Admissible data formats:
-XXXX.XX -XXXX.X -XXXX.
-XXXX
XXXX.XX XXXX.X
XXXX.
XXXX
-XXX.XX -XXX.X -XXX.
-XXX
XXX.XX XXX.X
XXX.
XXX
-XX.XX -XX.X -XX.
-XX
XX.XX XX.X
XX.
XX
-X.XX -X.X -X.
-X
X.XX X.X
X.
X
-.XX
-.X
.XX
.X
G.5.2 Interface read commands (data request from the thermostat)
Command
Meaning
IN_PV_00
Interrogation of bath temperature (outflow temperature)
IN_PV_01
Interrogation of controlled temperature(int./ext., Pt/ext., Analog/ ext. serial).
IN_PV_03
Interrogation of external temperature TE (Pt100).
IN_PV_04
Interrogation of external temperature TE (Analog Input).
IN_PV_10
Interrogation of bath temperature in 0.001 °C.
IN_PV_13
Interrogation of external temperature TE (Pt100) in 0.0001 °C.
IN_SP_00
Interrogation of temperature set value.
IN_SP_01
Interrogation of pump power level.
IN_SP_02
Interrogation of cooling mode (0 = OFF / 1 = ON / 2 = AUTOMATIC).
IN_SP_03
Interrogation of overtemperature switch-off point.
IN_SP_04
Interrogation of outflow temperature limit TiH.
IN_SP_05
Interrogation of outflow temperature limit TiL.
IN_PAR_00
Interrogation of the control parameter Xp.
IN_PAR_01
Interrogation of the control parameter Tn (181 = OFF).
IN_PAR_02
Interrogation of the control parameter Tv.
IN_PAR_03
Interrogation of the control parameter Td.
IN_PAR_04
Interrogation of the control parameter KpE.
IN_PAR_05
Interrogation of the control parameter TnE (response: XXXX; 9001 = OFF).
IN_PAR_06
Interrogation of the control parameter TvE (response: XXXX; 5 = OFF).
IN_PAR_07
Interrogation of the control parameter TdE (response: XXXX.X).
IN_PAR_09
Interrogation of the max. correcting quantity limit.
IN_PAR_10
Interrogation of the control parameter XpF.
IN_PAR_14
Interrogation of setpoint offset.
IN_PAR_15
Interrogation of the control parameter PropE.
IN_DI_01
Status of Contact Input 1: 0 = open/ 1 = closed.
IN_DI_02
Status of Contact Input 2: 0 = open/ 1 = closed.
IN_DI_03
Status of Contact Input 3: 0 = open/ 1 = closed.
IN_DO_01
Status of Contact Output 1:
0 = NO contact open/ 1 = NO contact closed.
IN_DO_02
Status of Contact Output 2:
0 = NO contact open/ 1 = NO contact closed.
IN_DO_03
Status of Contact Output 3:
0 = NO contact open/ 1 = NO contact closed.
IN_MODE_00
Keypad: 0 = released / 1 = locked.
IN_MODE_01
Control: 0 = int./ 1 = ext. Pt100 / 2 = ext. Analog / 3 = ext. Serial.