ECO GOLD
25/08/2011/ YACE0088
Technical data and graphs
67
10
Technical data and graphs
The figures were determined according to DIN 12876.
Data applicable to all ECO GOLD thermostats
Ambient temperature range
°C
5 ... 40
Relative humidity
Maximum relative humidity 80 % at 31 °C and decreasing linearly
to 50 % up to 40 °C.
Contamination level
2
Setting resolution
K
±0,01
Display resolution
K
±0,01
Temperature measurement
Absolute accuracy
K
±0,2
Temperature accuracy
K
±0,01
Pump type/number of power levels
Pressure pump/6
Discharge pressure, max.
bar
0.55
Discharge flow, max.
L/min
22
Viscosities of the heat carrier liquid
mm²/s
Heating range: maximum 150;
Control range:
≤
30
Display field
TFT display 3.5"; 320 x 240 pixel
Programmer
5 programs with a total of 150 temperature/time segments
Standard interface
USB
Class of protection
IP 21
Classification
lII
Labeling
FL
(
suitable for flammable and non-flammable liquids)
Overvoltages
Overvoltage Category II and transient overvoltages according to
Category II.
Class of protection for electrical
operating equipment DIN EN 61140
(VDE 0140-1)
Class I