ECO GOLD
25/08/2011/ YACE0088
Appendix with settings
87
Code 1XX Safety sys-
tem
Meaning
Code
2XX Command
Meaning
1
CAN receive overf
Overflow during CAN reception
1
CAN receive overf
Overflow during CAN reception
2
Watchdog Reset
Watchdog reset
2
Watchdog Reset
Watchdog reset
5
Heat 1 failed
Heater 1 defective
3
Clock Error
Battery fault
6
Heat 2 failed
Heater 2 defective
9
Unknown Modul
Unknown module connected
7
Invalid Parameter
Inadmissible parameter in memory
10
SW control too old
Software version of control panel too old
8
CAN system
Problem during internal data interchange
11
SW safety too old
Software version of protection too old
9
Unknown Modul
Unknown module connected
12
SW command too
old
Software version of command remote
control unit too old
10
SW control too old
Software version of control panel too old
13
SW cool too old
Software version of cooling module too old
11
SW safety too old
Software version of protection too old
14
SW analog too old
Software version of analog too old
12
SW command too
old
Software version of command remote
control unit too old
15
SW serial too old
Software version of RS232 too old
13
SW cool too old
Software version of cooling module too old
16
SW contact old
Software version of contact module too old
14
SW analog too old
Software version of analog too old
17
SW Valve 0 old
Software version of solenoid valve 0 too old
15
SW serial too old
Software version of RS232 too old
18
SW Valve 1 old
Software version of solenoid valve 1 too old
16
SW contact old
Software version of contact module too old
19
SW Valve 2 old
Software version of solenoid valve 2 too old
17
SW Valve 0 old
Software version of solenoid valve 0 too old
20
SW Valve 3 old
Software version of solenoid valve 3 too old
18
SW Valve 1 old
Software version of solenoid valve 1 too old
21
SW Valve 4 old
Software version of solenoid valve 4 too old
19
SW Valve 2 old
Software version of solenoid valve 2 too old
26
SW HTC old
Software version of high temperature cooler
too old
20
SW Valve 3 old
Software version of solenoid valve 3 too old
21
SW Valve 4 old
Software version of solenoid valve 4 too old
26
SW HTC old
Software version of high temperature cooler
too old
27
SW Ext Pt100 old
Software version of external Pt100 too old
Code 4XX Analog mod-
ule
Meaning
Code
5XX Serial
(RS232/485)
Meaning
1
CAN receive overf
Overflow during CAN reception
1
CAN receive overf
Overflow during CAN reception
2
Watchdog Reset
Watchdog reset
2
Watchdog Reset
Watchdog reset
9
Unknown Modul
Unknown module connected
9
Unknown Modul
Unknown module connected
10
SW control too old
Software version of control panel too old
10
SW Contr. too old
Software version of control panel too old
11
SW safety too old
Software version of protection too old
11
SW safety too old
Software version of protection too old
12
SW command too
old
Software version of command remote
control unit too old
12
SW command too
old
Software version of command remote
control unit too old
13
SW cool too old
Software version of cooling module too old
13
SW cool too old
Software version of cooling module too old
14
SW analog too old
Software version of analog too old
14
SW analog too old
Software version of analog too old
15
SW serial too old
Software version of RS232 too old
15
SW serial too old
Software version of RS232 too old
16
SW contact old
Software version of contact module too old
16
SW contact old
Software version of contact module too old
17
SW Valve 0 old
Software version of solenoid valve 0 too old
17
SW Valve 0 old
Software version of solenoid valve 0 too old
18
SW Valve 1 old
Software version of solenoid valve 1 too old
18
SW Valve 1 old
Software version of solenoid valve 1 too old
19
SW Valve 2 old
Software version of solenoid valve 2 too old
19
SW Valve 2 old
Software version of solenoid valve 2 too old
20
SW Valve 3 old
Software version of solenoid valve 3 too old
20
SW Valve 3 old
Software version of solenoid valve 3 too old
21
SW Valve 4 old
Software version of solenoid valve 4 too old
21
SW Valve 4 old
Software version of solenoid valve 4 too old
26
SW HTC old
Software version of high temperature cooler
too old
26
SW HTC old
Software version of high temperature cooler
too old
27
SW Ext Pt100 old
Software version of external Pt100 too old
27
SW Ext Pt100 old
Software version of external Pt100 too old