87
Service A: Professional Series Distilled Water Cooling
System
Tools-Supplied or Required:
Phillips Screw Driver #2
Metric Hex Key Wrench Set (Metric 1.5 to 6 mm set)
Adjustable Wrench (1 1/8” range) or Open End (1.0”)
Distilled Water
Distilled Water Bottle Siphon, manual water pump, or Battery or Basting
Fill Bulb
Absorbent Paper Towels (Ex.-Scott Shop Towels/Blue)
Clean Room Vinyl/PVC Gloves (
Powder-Free
/DEHP & DOP Free)
Clean plastic container / bucket (≥6 Liters / 1.5 gallons)
Small Funnel
ATTENTION:
The system must be filled with
distilled water before turning ON the system.
CAUTION: Only use distilled water from a
reputable supplier. Use of tap water will
damage the device and void the warranty.
(
Note 1:
iWeld Professional Series-The Distilled Water
Bottle can be filled; water filter replaced & bottle removed
from the left side of the machine.)
(Note 2:
The Distilled Water Bottle does not need to be
removed for the initial filling of Distilled Water or topping
off the Distilled Water.
Summary of Contents for iWeld 900 Series
Page 2: ...LaserStar Workstation Operation Maintenance Manual 900 Series Ver 991 6 November 2019 2...
Page 134: ...LaserStar Workstation Operation Maintenance Manual 900 Series Ver 991 6 November 2019 134...
Page 136: ...LaserStar Workstation Operation Maintenance Manual 900 Series Ver 991 6 November 2019 136...
Page 148: ...LaserStar Workstation Operation Maintenance Manual 900 Series Ver 991 6 November 2019 148...
Page 150: ......
Page 152: ...LaserStar Workstation Operation Maintenance Manual 900 Series Ver 991 6 November 2019 150...
Page 153: ......
Page 154: ......
Page 155: ......
Page 156: ......
Page 157: ......
Page 158: ......
Page 159: ......
Page 160: ......