The following explains each of the display’s components:
Display label - Shows the current display (for example,
“Ln-a”).
Percentage settings are entered with
a 1/100th dB resolution, but are dis-
played to the nearest 1/10th dB when
10 dB or greater.
Percentile values & levels - The L
n
(percentile levels) level
is that level which is exceeded ‘n’ percent of the measure-
ment time. By modifying the six settings you can select any
six percentile values to be displayed at a time. For example,
if you set one of these settings to 10.5, then L
10.5
would be
one of the six percentiles that is displayed. The value of
L
10.5
is the level that has been exceeded 10.5% of the time
during the overall measurement.
Ln-b Display
The Ln-b display consists of five major components:
•
Display label
•
Current % record
•
Ln
•
Elapsed Time
•
Ln graph
The
Ln Centiles-b display is shown below for the LOG
instrument:
The following explains each of the display’s components:
Display label - Shows the current display (for example,
“Ln-b Centiles”).
Display label
Ln graph
Ln
ElapsedTime
Current%
Summary of Contents for System 824
Page 4: ...824 Reference Manual iv...
Page 18: ...824 Reference Manual xiv...
Page 34: ...1 16 824 Reference Manual 11 8 00...
Page 88: ...3 38 824 Reference Manual 11 8 00...
Page 118: ...5 10 824 Reference Manual 9 12 00...
Page 180: ...6 62 824 Reference Manual 11 8 00...
Page 314: ...9 36 824 Reference Manual 11 8 00...
Page 356: ...10 42 824 Reference Manual 11 8 00...
Page 371: ...11 2 00 FFT Optional 11 15...
Page 372: ...11 16 824 Reference Manual 11 2 00...
Page 388: ...12 16 824 Reference Manual 11 8 00...
Page 398: ...A 10 824 Reference Manual 9 12 00...
Page 485: ...9 13 00 C 17 Windscreen Corrections...
Page 503: ...9 13 00 C 35 Graph 20 824 in the YZ plane fitted with the 2541 micro phone 12 5 kHz...
Page 504: ...C 36 824 Reference Manual 9 13 00...
Page 536: ...824 Reference Manual x...