11/8/00
AUD (Optional)
12-11
You will notice that the Fundamental - reference level is
now zero as the reference level was set to equal the funda-
mental Level. The Fundamental level still shows the dB
level of the incoming signal.
Linearity information is only avail-
able on screen at the time of the
measurement and is not stored with a
file or printed to a report.
By varying the amplitude level you will see the Fundamen-
tal - reference level change. Usually audiometers have steps
of 5 or 10 dB. You can also see the effect frequency has on
the accuracy of the amplitude linearity by varying the fre-
quency.
Pulse/FM Mode
In the Pulse/FM mode, the 824 can be used for the following
measurements:
•
pulse measurements
•
frequency modulation measurements
This mode consists of two screens: the Pulse/FM-a and
Pulse/FM-b screens.
Pulse/FM-a Screen
To access the Pulse/FM-a screen, press the VIEW key
,
then select Pulse/FM with the up/down arrows and press the
check key
.
This will display the Pulse/FM-a screen.
Summary of Contents for System 824
Page 4: ...824 Reference Manual iv...
Page 18: ...824 Reference Manual xiv...
Page 34: ...1 16 824 Reference Manual 11 8 00...
Page 88: ...3 38 824 Reference Manual 11 8 00...
Page 118: ...5 10 824 Reference Manual 9 12 00...
Page 180: ...6 62 824 Reference Manual 11 8 00...
Page 314: ...9 36 824 Reference Manual 11 8 00...
Page 356: ...10 42 824 Reference Manual 11 8 00...
Page 371: ...11 2 00 FFT Optional 11 15...
Page 372: ...11 16 824 Reference Manual 11 2 00...
Page 388: ...12 16 824 Reference Manual 11 8 00...
Page 398: ...A 10 824 Reference Manual 9 12 00...
Page 485: ...9 13 00 C 17 Windscreen Corrections...
Page 503: ...9 13 00 C 35 Graph 20 824 in the YZ plane fitted with the 2541 micro phone 12 5 kHz...
Page 504: ...C 36 824 Reference Manual 9 13 00...
Page 536: ...824 Reference Manual x...