Spartan 730/730IS Dosimeter Manual
A-15
The Spartan 730 dosimeter complies with ANSI S1.25-2007 as specified below.
A.2.4 FCC/IC Compliance
Contains
FCC ID: 2AA9B04
This device complies with part 15 of the FCC rules.
Operation is subject to the following two conditions. (1) This device may not cause harmful interference. (2) This device must
accept any interference received, including interference that may cause undesired operation.
Contains
IC: 12208A-04
This device complies with Industry Canada license-exempt RSS standard(s). Operation is subject to the following two
conditions: (1) this device may not cause interference, and (2) this device must accept any interference, including interference
that may cause undesired operation of the device.
Le présent appareil est conforme aux CNR d'Industrie Canada applicables aux appareils radio exempts de licence.
L'exploitation est autorisée aux deux conditions suivantes: (1) l'appareil ne doit pas produire de brouillage, et (2) l'utilisateur
de l'appareil doit accepter tout brouillage radioélectrique subi, même si le brouillage est susceptible d'en compromettre le
onctionnement.
Class Designation
Class 2 with variable A or C frequency weighting, variable for F or S time constant,
variable Criterion Level, Threshold, and Exchange Rate.
Upper Limit Indicator
Overload is indicated when the upper limit has been exceeded and held for 4
seconds.
Operating Range
See
.
Pulse Range
88 dB
Microphone Frequency
Response
The frequency response is listed in section A.2.1 Microphone, Frequency Response,
and illustrated graphically in section
.
Maximum Safe SPL
143 dB peak
Sensitivity to Magnetic Field
No sensitivity detected at 80A/m
Temperature and Humidity
Range
See
Environmental Conditions on page A-5
. This also applies to storage
conditions.
Calibration Procedure
See
2.6.1 Calibrating Your Spartan Dosimeter on page 26
.
Mounting on Wearer
See
Deploying the Dosimeter on page 4-53
Test Points
The means to insert electrical signals is provided by using the ADP106 (Adapter
input BNC to 1/4-inch Mic 12 pF) The electrical impedance of the source connected
to the ADP106 shall be 600
Ω
or less.
Maximum Linear Peak Level
140 dB peak
Recalibration Period
1 year