Spartan 730/730IS Dosimeter Manual
A-12
A.1.7 Event Sound Recording Specifications
The following specifications apply to the Spartan 730 dosimeter option 730-ESR.
Occurrence Date &
Time Stamp
yyyy-mm-dd hh:mm:ss
Time History Metrics
Other Indicators
Overload, Motion, Bump, Run, Stop, Pause
File Format
16-bit WAV
Sample Rate
8000 sps
Recording Time
2 second pretrigger
+10 second record time
12 seconds per sound recording total
Minimum Time Between
Recordings
Selectable (15—3600 sec) default is 30 sec.
Trigger Source
Selectable (40–140 dB) default is 85 dB
Label
Units
Equivalent Level A-weight (t)
L
Aeq
dB
Equivalent Level C-weight (t)
L
Ceq
dB
Maximum Level A-weight Slow (t)
L
ASmax
dB
Maximum Level A-weight Fast (t)
L
AFmax
dB
Peak Level C-weight (t)
L
Cpkmax
dB
Peak Level Z-weight (t)
L
Zpkmax
dB
Time Weighted Average (Q3) (t)
TWA3
dB
Time Weighted Average (Q5) (t)
TWA5
dB