Spartan 730/730IS Dosimeter Manual
Defining a Measurement Configuration
35
this setting, exposure levels are recorded in terms of a table of metrics tracked
every second (or minute, respectively).
About the Octave Band Analysis Option (730-OB1)
The Octave Band Analysis (OBA) feature on your Spartan 730 allows you to
analyze the measured noise exposure in full octaves.
Industrial Hygiene professionals sometimes use octave band analysis when
measuring personal noise exposure. Octave bands allows you to see the
frequency composition of a sound. Among other uses, studying the frequency
composition of measured noise can help you determine the most appropriate
type of hearing protection for the specific noise frequencies shown.
About the Remove Bluetooth Capability Option (730-NBT)
This firmware option allows you to disable the dosimeter’s built-in Bluetooth
capability, so that this dosimeter model can be used where radios are not
allowed. When enabled, the meter can be set up, controlled, and data
downloaded by using the included USB cable.
Enabling the Option On the Meter
Before you begin:
•
Power on the dosimeter.
•
Stop any measurement in progress
Step 1.
Press and hold
select
to open the dosimeter Menu.
Step 2.
Highlight, then
select
Settings
to enter the Settings menu.
Step 3.
Highlight, then
select
Meter
. This opens the Meter menu.
In this section, use
UP
or
DOWN
on the dosimeter to
highlight an option. Use
SELECT
to set the option.