![Larson Davis Spartan 730 Reference Manual Download Page 64](http://html1.mh-extra.com/html/larson-davis/spartan-730/spartan-730_reference-manual_3719581064.webp)
Spartan 730/730IS Dosimeter Manual
Setting Event Sound Recording Options
62
Part 1:
Setting Options for Event Recording
This section reviews how to set the event sound recording options, which include
the following values: Trigger Source, Trigger Level, and Minimum Interval.
Step 1.
From the Evt Sound Rec menu, highlight, then
select
Trigger Src
. This
opens the Trigger Src screen.
Step 2.
Press
down
to highlight an available option, then press
select
to set the
option and return to the Event Sound Rec menu.
Step 3.
From the Evt Sound Rec menu, highlight, then
select
Trigger Levels
.
This opens the Trigger Levels screen.
Step 4.
Enter the
Trigger Level
in dB.
Setting the Trigger Level by Using the Dosimeter Buttons
•
Press
up
or
down
to change the digits.
•
Press
select
to set the digit and progress to the right.
•
Set all digits to exit the Trigger Level screen.
•
Quick press
power
to step back to the left at any time.
Step 5.
From the Evt Sound Rec menu, highlight, then
select
Min Interval
. This
opens the Min Interval screen.
Step 6.
Set the
Min Interval
in seconds using the dosimeter buttons as shown
in step 4.
In this section, use
UP
or
DOWN
on the dosimeter to
highlight an option. Use
SELECT
to set the option. Quick
press
power
to step up one level in a menu.
Available options include LCS, LAF, LAeq1s, LCeq1s, and LAS.
Minimum Interval is the number of seconds between event
recordings. The default setting is 30 seconds. Set a longer Min
Interval to minimize the number of recorded events and
conserve dosimeter memory.