Spartan 730/730IS Dosimeter Manual
Defining a Measurement Configuration
32
Before you begin:
•
Power on your dosimeter.
•
If applicable, stop any measurement in progress.
•
Step 1.
Press and hold
select
to open the dosimeter Menu.
Step 2.
Highlight, then
select
Settings
to enter the Settings menu.
Step 3.
Highlight, then
select
Dosimeter 1
. This opens the
Dosimeter 1
menu.
Step 4.
Highlight, then
select
Configure
. This opens the Select Config screen.
Step 5.
Highlight, then
select
a configuration from the available options also
shown in
TAKE NOTE
The following table shows properties of the international standard
configurations for virtual dosimeters 1, 2, 3, and 4. The Custom
configuration is also an option for each dosimeter.
Table 3.1 Properties of International Standard and Custom Configurations
In this section, use
up
or
down
to highlight an option.
Use
select
to set the option. Press
power
to step up
one level in a menu.
OSHA-PEL
OSHA-HC
ACGIH
NIOSH
Custom
Mode
Dose
Dose
Dose
Dose
Dose or ISO
Exchange Rate
5
5
3
3
3, 4, 5, or 6
Criterion Level
90.0
85.0
85.0
85.0
User selected
Threshold Enable
Enabled
Enabled
Enabled
Enabled
User selected
Threshold
90.0
80.0
80.0
80.0
User selected
Shift Time
8.0
8.0
8.0
8.0
User selected
Frequency Weight
A-Weight
A-Weight
A-Weight
A-Weight
User selected
Detector
Slow
Slow
Slow
Slow
Slow
Peak Frequency Weight
C-Weight
C-Weight
C-Weight
C-Weight
User selected