6.7 Troubleshooting Dosimeter Power
When the Spartan dosimeter does not power on, especially after being stored for
prolonged periods, do the following:
Step 1.
Connect the meter to a USB power source, and do not press any buttons
while connecting power.
Step 2.
Wait at least one minute, then press the power button for 1 second to
attempt to power the meter. If the meter turns on, skip to step 6.
Step 3.
Repeat step 2 up to three times until the meter powers on. If it still does
not power on, move to step 4.
Step 4.
Press and hold the power button for 15 seconds to force the meter to
power off.
Step 5.
Then, press the power button for 1 second to attempt to power the
meter. If the meter powers on, move to step 6. If the meter does not
power on, contact Larson Davis to begin the RMA/repair process.
.
Step 6.
Charge the meter.
• When preparing to deploy, charge the battery to 100% (or as full as
possible in the given time).
• When preparing to store, charge to 50%, then disconnect and
power down the meter.