34
CHAPTER 4:
Computer Interface Operation
RANGE?
Current Range Query
Input
RANGE?[term]
Format
nn
Returned
<numeric key value>[term]
SETI
Set User Current Command
Input
SETI
Format
±nnne-nn
<user current value>
Current value as a three-
digit number including a
decimal place
Remarks
User current is settable between 100 nA and 100
mA. The sign of the user current value determines
the polarity of the output. The IPOL command does
not affect this setting.
SETI?
Set User Current Query
Input
SETI?
Format
±nnn.nnne-nn
Returned
<user current value>
Remarks
User current will be returned in engineering nota-
tion with three significant digits.
SETSAVE
Power on Setup State
Input
SETSAVE
Remarks
The command saves the present range, polarity,
and user current value. This saved state will be
loaded on future power ups.
Summary of Contents for 121
Page 6: ......
Page 10: ...iv ...
Page 46: ...36 CHAPTER 5 Options and Accessories ...
Page 54: ...44 CHAPTER 6 Service ...
Page 60: ...50 APPENDICES B 1 1 CE Declaration for Model 121 ...
Page 61: ...51 B 1 2 CE Declaration for the Power Supply Part Number 109 132 ...
Page 62: ...52 APPENDICES ...