31
IDN?
Identification Query
Input
IDN?[term]
Format
s[4],s[8],s[7],n.n
<manufacturer>
Manufacturer ID
<model>
Instrument model number
<instrument serial>
Instrument serial number
<firmware version>
Instrument firmware
version
Example
LSCI,MODEL121,1234567,1.0
RST
Reset Instrument Command
Input
RST[term]
Remarks
Sets instrument parameters to power-up settings.
BRIGT
Display Brightness Command
Input
BRIGT<display contrast>[term]
Format
nn
<display brightness> 0–15
Remarks
Sets the display contrast for the front panel seven-
segment display. A higher number makes the dis-
play brighter. The default setting is 8. The display
can be turned off with setting the brightness to 0.
BRIGT?
Display Brightness Query
Input
BRIGT?[term]
Format
nn (refer to command for description
Returned
<display brightness>[term]
COMP?
Compliance Limit Query
Input
COMP?[term]
Format
n
Returned
<compliance status> 0 = normal operation, 1 = in
compliance limit
Remarks
Returns the voltage compliance limit status of the
current source output