28
CHAPTER 4:
Computer Interface Operation
D
Properly format and transmit the query including the terminator as
one string
D
Prepare to receive a response immediately
D
Receive the entire response from the instrument including the ter-
minator
D
Guarantee that no other communication is started during the
response or for 50 ms after it completes
D
Not initiate communication more than 20 times per second
Failure to follow these simple rules will result in inability to establish
communication with the instrument or intermittent failures in commu-
nication