S
YSTEM
M
AINTENANCE
WHEN TO CHANGE THE 0.2 MICRON FINAL FILTER (FL0016-1)
The final filter is able to remove particulates down to 0.2 µm to provide bacteria-free water .
Page 48
Issue
Solutions
When the flow rate through the final filter drops below
one liter per minute.
Replace the 0.2 Micron Final
Filter .
When product water continues to flow from the 0.2 mi-
cron final filter after the Dispense button has been re-
leased.
When the bacteria levels increase above the recom-
mended ASTM Standards of Bacterial count1 0/1000mL
Table 6.5
LT0033-4