Page 27
G
RAPHICAL
T
OUCH
S
CREEN
Settings Screen
LabStrong™ Cop.
Ref. # Description
Function
67
Edit Date &
Time
Use this setting to change
the date and time of the
unit.
68
Sound
Settings
Use this setting to adjust the
volume of the unit.
69
Exhaustion
Threshold
The
Exhaustion Threshold
is
the temperature-
compensated resistivity val-
ue below which cartridges
will be considered exhaust-
ed. Further explanation can
be found in the Maintenance
section.
70
Reset to Fac-
tory Defaults
Setting should only be used
if the unit needs to be reset
back to default values. Once
you press ‘reset’, the action
can not be undone.
71
Temperature
Compensa-
tion
While in ‘ON’ the resistivity
will be compensated to
25°C. In ‘OFF’ the resistivity
reading will not be
compensated to 25°C.
67
68
69
70
71
Figure 3.20
Table 3.21