LabSmith HVS448 User Manual Download Page 8

 

Master Interlock Input 

The Interlock inpu

t

 is the center terminal of a BNC connector located on the back panel of the HVS448.  The interlock signal is 

TTL-compatible (0–5 V) and weakly pulled to 5 V via an ~20 kΩ internal resistor. When the interlock signal is in its high state (>0.8 
V), the high voltage connectors are disconnected from the power supply by relays, and the internal high voltage supply is turned 
off.  Because of the internal pull-up, a high state can be maintained by leaving the signal open

.  

  

High voltage is enabled by maintaining a TTL low signal (<0.2 V) to the interlock.  This can be accomplished by

 

connecting the 

supplied 50-Ω terminator (

Figure 2

) to the interlock connection

,

 connecting to an interlock switch to ground, or connecting to a 

digital interlock signal that is active low. 

To test if the interlock is satisfied (low): on the front panel click the 

Enable

 then 

All

 buttons.  If the interlock is satisfied this will 

enable the high voltage and turn on a green light over the 

All

 button.  If the light is red, the interlock is not satisfied. 

 

Figure 2. 50-Ω

    

BNC Terminator for Interlock 

Disable/Enable High Voltage 

Disabling high voltage is a quick way to place the system in a “safe” state.  The high voltage connectors on the back panel will 
float to the driving voltage when disabled, regardless of the sequence or the states of other channels.  High voltage can be 
enabled or disabled from the software or using front panel controls. This setting is nonvolatile and is remembered on power-up

Disable High Voltage: 

 

From the front panel, push 

Disable, t

hen

 All; 

 

or, from the software click the

 Disable High Voltage

 

 toolbar button or , choose 

Actions>Disable High Voltage

.

 

When the high voltage is disabled the LED over the ALL button will turn off, or will glow red if the Interlock is not satisfied. 

Enable High Voltage: 

 

From the front panel, push 

Enable, t

hen

 All; 

 

or, from the software, click the

 Enable High Voltage

 

 toolbar button or choose 

Actions>Enable High Voltage

.

 

When the high voltage is enabled the LED over the ALL button will glow green if the interlock is satisfied or red if the Interlock is 
not satisfied. 

 

High Voltage Connections 

Each channel of the HVS448 includes a single high voltage connector. Each channel can function as a voltage/current source or 
monitor, so this connector acts as either an input or output, depending on the function of the channel at any time.  

The total system output (all channels considered) will not exceed the maximum unit current (see 

Table 1

). The system will 

automatically shut down the high voltage supplies if this current is exceeded. Each channel’s maximum and minimum voltage 
and current can be limited in software to protect external devices and equipment (see 

Channel Output Limits

). Under no 

condition will the voltage or current at the channel connection exceed these software limits. 

WARNING:  

DO NOT connect or disconnect cables from an enabled high voltage channel. Doing so could result in 
personal injury and/or damage to the HVS448 or external equipment. You must disable the high 
voltage connectors or turn off the HVS448 before plugging or removing equipment. Use the Disable 

HV button 

 on the toolbar or select Actions> Disable High Voltage from the dropdown menu tab. 

WARNING:  

Do NOT connect a high voltage channel to a digital input or output. While the digital inputs and 
outputs are protected to 115 VAC, connecting these different voltage levels could damage the HVS448 
and/or any connected equipment. 

Summary of Contents for HVS448

Page 1: ...LabSmith HVS448 High Voltage Sequencer User Guide Version 1 5 For operation and control of LabSmith HVS448 High Voltage Sequencer and Sequence Driver Software...

Page 2: ...duced or distributed without the consent of LabSmith LabSmith HVS448 and Sequence are trademarks of LabSmith Inc Microsoft Windows NT XP 7 and 8 are registered trademarks of Microsoft Corporation Adob...

Page 3: ...ions 14 Digital Connections 14 Recalibrating the Channels 15 CHAPTER 4 SEQUENCE SOFTWARE OVERVIEW 16 Field Upgradeable Firmware 16 Sequence Files 16 Online and Offline Operation 16 Sequence Software E...

Page 4: ...30 CHAPTER 7 CHANNEL PROGRAM INSTRUCTIONS 31 Nomenclature 31 Execution Controls 31 Inter channel Signaling and Triggering 33 Channel Functions 35 Register Functions 36 Calculator Functions 37 Reverse...

Page 5: ...igh Voltage Sequence Available Models Model Max Output Voltage V Max Voltage Differential V Max Output Current mA Max Output Current Per Channel mA Current Monitor Resolution nA Voltage Monitor Resolu...

Page 6: ...ting Help This guide is your main source for information on operating the HVS448 and Sequence software Choose Help About Sequence then click the Open User Manual link to browse this manual electronica...

Page 7: ...ttaching any one of the grounding leads to a ground source WARNING Do not handle cable terminations or clips while they are connected to an energized circuit The micro and nano clips do not protect fr...

Page 8: ...ton or choose Actions Disable High Voltage When the high voltage is disabled the LED over the ALL button will turn off or will glow red if the Interlock is not satisfied Enable High Voltage From the f...

Page 9: ...Using the Sequence Wizards to Create a Sequence File Sequence Wizards simplify the process of creating sequences by letting you enter voltages and currents directly for each step of the sequence A se...

Page 10: ...TTL signal at any of the four inputs In1 In4 5 Enter the setting changes for each channel during this step a Select Unchanged to leave the channel in the same state as the previous sequence step b Se...

Page 11: ...aveform shape you wish to apply then set Offset V the DC offset if any applied to the waveform Amplitude V the peak to peak voltage differential of the waveform Note By default generated waveforms are...

Page 12: ...VS448 now box b If the HVS448 was online prior to programming choose Actions Program All 5 In the left pane click Sequence to open the Sequencer Control Panel 6 Press A at the bottom of the Sequencer...

Page 13: ...ply that function to The function button will remain lit after you select a channel indicating that you can select additional channels Push the function button again to deselect it Selecting a differe...

Page 14: ...button glows a steady red and all buttons are dark The front panel can be unlocked by choosing Actions Unlock Front Panel This is the default condition When the panel is unlocked the power LED glows g...

Page 15: ...st 30 minutes before calibration 4 Run the Sequence software 5 In Sequence enter Online mode by clicking the toolbar button 6 Click the toolbar button to enable high voltage The All indicator on the H...

Page 16: ...rnal source is supplied then they will indicate near 0 V When you open a Sequence file the channels remain inputs Click the Update All button at the bottom of Sequence s right pane to set all of the c...

Page 17: ...asurement interval is the frequency with which measurement values are updated in milliseconds The measurement interval must be an integer in the range of 1 to 1000000000 The new measurement interval w...

Page 18: ...nd values continuously Note The default monitor setting is governed by the Monitor Settings dialog To set the default choose View Options Monitoring see Monitor Settings A channel can monitor either i...

Page 19: ...rent settings If Monitoring is turned on for a channel any problems on that channel will cause a message to appear in the Errors field Click the Trip Setting button to choose the error responses for a...

Page 20: ...current experiment The voltage and current sliders for each channel will adjust to reflect your choice Update All synchronizes the data from all channels reads the calibration data updates all channel...

Page 21: ...s settings expand the channel in the left pane Figure 17 Figure 17 Channel A s High Voltage Settings Shown in the Left Pane The function and limits from the Channel Control Panels will be displayed I...

Page 22: ...h let a particular channel respond to a combination of internal or external events Manual Changes During an Automated Sequence When you run a sequence the Sequence software will automatically change c...

Page 23: ...r step instructions until the master sequencer terminates the program by issuing its own Exit instruction see Exit If any special shutdown procedures are required by a channel they should be run in St...

Page 24: ...s not possible with flags reducing the potential for software bugs For example the earlier example of coordinating channels to make complementary square waves may be better implemented using triggers...

Page 25: ...cer Control Panel click Sequencer in the left pane Figure 18 Many of these controls will be familiar from the Manual Power Supply Control Panel Check the Monitor box next to any channel to view the pr...

Page 26: ...esh the values Check the Continuously Monitor Measurements box to update the values on a continual basis The Navigate Programs drop down menu shows the channel programs Select a program to view its co...

Page 27: ...me hierarchy as the software sequences consist of steps and programs programs consist of instructions In the left pane double click Sequencer to expand and view the master and channels controls Figure...

Page 28: ...diting a Program Instruction in the Left Pane To add a new instruction right click on an instruction within that program From the drop down menu that appears choose New Instruction A new instruction w...

Page 29: ...lbar button Figure 28 Programming Keypad To use the keypad open the text of a sequence or program in the right pane Place the cursor where you want to add a command then click the required instruction...

Page 30: ...struction will also appear in red as will the sequence containing the program and the Sequence icon itself to help you quickly spot the problem In the example in Figure 30 the same label appears twice...

Page 31: ...All program labels that start with a capital letter are globally known by all programs of the particular channel No two global labels can be the same for any of the programs of a channel Local Labels...

Page 32: ...ns between these timing commands provided it is less than the specified duration Example Do something that takes 12 5 ms 12 5 ms Pause pause until the pause counter reaches 12 5 ms execute instruction...

Page 33: ...p instruction is Continue the program will remain stopped Example Stop The program waits here for a Step change or user command Exit Usage Exit The Exit instruction takes no argument and behaves diffe...

Page 34: ...F2 F3 F4 This instruction tests whether the specified flag is clear If it is clear the next instruction is executed If it is not clear the next instruction is skipped This instruction has no other ef...

Page 35: ...cuted if it is not clear the next instruction is skipped This instruction has no other effect Example Label a InputClear Loop while the trigger input is clear Goto a Get here when the trigger input is...

Page 36: ...in RAM memory to 13 registers A M Unlike the stack which can store up to 16 numbers sequentially the registers can be accessed randomly The registers are cleared set to zeros at system startup but ar...

Page 37: ...ains 99 The X register now contains 117 The behavior of each operation on the numerical stack is detailed for each instruction Integral Representation of Real Numbers All numbers in the stack are 24 b...

Page 38: ...uA X contains 16 uA Y contains 10 s SwapXY X contains 10 s Y contains 16 uA ClearX Usage ClearX ClearX sets the X register to 0 without changing the other values in the stack Example 10 X contains 10...

Page 39: ...6 X contains 166 Y contains 10 12 X contains 12 Y contains 166 X contains 13 166 12 13 833 Y contains 10 Mod Usage Mod The instruction Mod removes the top two numbers on the stack X and Y and places t...

Page 40: ...th 0 X 0 X is less than 0 X 0 X is less than or equal to 0 X 0 X is equal to 0 X 0 X is not equal to 0 X 0 X is greater than or equal to 0 X 0 X is greater than 0 If the test is true the next instruct...

Page 41: ...41 10 X contains 10 16 X contains 16 Y contains 10 X Y False Goto Initialize Does not execute this instruction X Y True Goto Inject Executes this instruction...

Page 42: ...ce All channel states remain as they are This is equivalent to the Pause Resume button on the HVS448 front panel Resume restarts progress of the current sequence This is equivalent to the Pause Resume...

Page 43: ...programming Paste lets you paste text within your programming Help lets you access Sequence s manual Enable high voltage output allows the high voltage connectors to respond to the current settings a...

Page 44: ...BNC B OUTA BNC HVC SHV BNC G HVE SHV BNC HVD SHV BNC IN1 BNC Sequencer Channel F 1 4 Input Output High Voltage Step 1 8 OUTC BNC HVH SHV BNC Programmable Trigger Logic Inputs Outputs in1 in2 in3 in4...

Page 45: ...e In the first step LOAD the voltages will control a flow from reservoir A to reservoir C Voltages applied to B and D will be used to pinch the flow at the intersection and define a time independent s...

Page 46: ...p A Loading 7 Click on the Step B tab to view its settings 8 Change the Step B name to Injection 9 Set the Switch Step setting to Only Manually 10 Set Channels A D to regulate D C voltages according t...

Page 47: ...g voltage sequence On the Manual Sequence Control Panel you should now see channels A D maintaining the voltages from 3 Table 4 If you are viewing the microchannel on a microscope you should observe a...

Page 48: ...p 9 Noisy Channels Operating at low voltages 200 V The HVS is designed for optimal performance at high voltages Use a higher common mode voltage to achieve the desired potential difference i e apply 1...

Reviews: