5
LABORIE Urocap IV Light Service Manual UCL-SM01 V7.00
INTRODUCTION
This manual is intended for servicing
LABORIE’s
Urocap
™
IV Light System that
collects data from the Uroflow transducer. The information located in this
manual is intended for the biomedical departments of hospitals or clinics.
SPECIFICATIONS AND SYSTEM INFORMATION
Upon request, LABORIE will make available any circuit diagrams, component
part lists, and other technical documentation directly related to the Urocap IV
Light system.
Table 1: Specifications
Transportation
and Storage
Conditions
Temperature: -10ºC to +50ºC (14ºF to 122ºF)
Humidity: 20% to 80% relative humidity (non-condensing)
Pressure: 700 hPa to 1014 hPa
Operating
Conditions
Temperature: +10ºC to +40ºC (50ºF to 104ºF)
Humidity: 30% to 75% relative humidity (non-condensing)
Pressure: 70 to 106 kPa
Urocap IV
(TRA851)
Size: 16.0 cm (6.3
”
) L X 16.0 cm (6.3
”
) W X 6.6 cm (2.6
”
) H
Weight: 0.3 kg (0.7 lbs)
Flow Range: 0 to 50 mL/s; Volume Range: 0 to 1500 mL Output
Channels: Flow; Volume
Sampling Rate: Flow = 20 Hz; Volume = 20 Hz
Charger: Connect to SL Power Electronics Corp. Model:
MW172KB054OF02 PSU only.
1