LABORIE UROCAP IV Service Manual Download Page 16

12 

LABORIE Urocap IV Light Service Manual UCL-SM01 V7.00 

 

Table 2: Troubleshooting - Mobile Device 

 

S

YMPTOM

(

S

P

OSSIBLE 

C

AUSE

(

S

C

HECK 

/ C

ORRECTIVE 

A

CTION

(

S

Urocap IV Light 
(UDS Light) 
software does not 
automatically 
appear when the 
device is turned 
on? 

Is 

Home 

application 
setting correct? 

1. 

Select the App 
Drawer icon on the 
Home screen. 

 

2. 

Scroll through the 
apps and select the 
Settings icon. 

 

3. 

Select the 

Home 

menu item. 

 

4. 

Select the 

Urocap IV Light (UDS 

Light) 

application. 

 

 

Summary of Contents for UROCAP IV

Page 1: ...LABORIE Urocap IV Light Service Manual UCL SM01 V7 00 UROCAP IV LIGHT Service Manual...

Page 2: ...oth is a registered trademark of Bluetooth SIG Android is a trademark of Google Inc Wi Fi is a registered trademark of the Wi Fi Alliance MOTOROLA and product images are trademarks or registered trade...

Page 3: ...Directives And Standards 4 Equipment Setup 5 Calibration 6 Cleaning And Maintenance 7 Battery Charging And Preventive Maintenance 8 Charging The Battery 8 Preventive Maintenance 9 Treating And Dispos...

Page 4: ...roubleshooting Bluetooth Connections 14 Troubleshooting Printer 14 Troubleshooting Urocap IV 15 Troubleshooting User Notification Messages 17 IEC 60601 1 2 2007 Table1 Requirements 22 IEC 60601 1 2 20...

Page 5: ...tlyrelated to the Urocap IV Lightsystem Table 1 Specifications Transportation and Storage Conditions Temperature 10 C to 50 C 14 F to 122 F Humidity 20 to 80 relative humidity non condensing Pressure...

Page 6: ...ode minimum 24 hours with fully charged battery Continuous run Minimum of 3 hours with fully charged battery Mobile Device Mount Metal Bracket Size 96 1 mm 3 78 H X 102 mm 4 00 W X 133 mm 5 22 Depth W...

Page 7: ...rnally powered equipment Classification of InstallationandUse Portable ClassI viaPSU InternallyPowered Equipment Battery Mode Degree of Protection TheUrocapIVuroflowmeterenclosureisclassified IP54 acc...

Page 8: ...4 2 ISO13485 EN61000 4 3 EN ISO13485 EN61000 4 4 EN ISO14971 EN61000 4 5 EN60601 1 EN61000 4 6 CAN CSAC22 2No 601 1 M90 EN61000 4 8 CAN CSAC22 2No 60601 1 EN61000 4 11 EN60601 1 6 CISPR11 EN62366 EN62...

Page 9: ...IV A the commode B the mobile device C and the printer D as shown Keepthe distance between the Urocap IV and the mobile device and the distance between the mobile device to wireless printer to less t...

Page 10: ...t or by pouring water into a beaker 1 Plug the Urocap IV into the charger and plug the charger into an outlet if the battery is not fullycharged 2 Tap Device Options then tapCalibration 3 Tap the cali...

Page 11: ...for ingress of water This means that the enclosure of the Uroflowmeter is able to handle splashes of water and liquid fromany direction but it is not protected against total immersion into water or a...

Page 12: ...repairs Check the transducer calibration of the Urocap IV every 6 months or whenever you suspect the transducer is offcalibration Pay close attention to the LED lights on the device If they indicate a...

Page 13: ...battery If misuse abuse or damage is suspected or any form of mechanical damage to the casing is visible then discontinue use and return the battery to LABORIE If a sudden change in the battery s abil...

Page 14: ...collection and recyclingservices ENVIRONMENTAL CONSIDERATION OF WASTE DISPOSAL Due to the fact that the Urocap IV is designed to perform Uroflow studies it is important to dispose of waste such as uri...

Page 15: ...t times it may requires pressing and holding for up to 30 seconds Is the Device out of power Is the battery drained Plug in the device and allow it to charge Mobile Device does not show the study scre...

Page 16: ...CORRECTIVE ACTION S Urocap IV Light UDS Light software does not automatically appear when the device is turned on Is Home application setting correct 1 Select the App Drawer icon on the Home screen 2...

Page 17: ...IBLE CAUSE S CHECK CORRECTIVE ACTION S 5 Click the Home button on the mobile device to set the Urocap IV Light UDS Light application as the Home application The Urocap IV Light UDS Light Home screen w...

Page 18: ...Device System settings Table 4 Troubleshooting Printer SYMPTOM S POSSIBLE CAUSE S CHECK CORRECTIVE ACTION S Printer power LED does not light up Printer power cable unplugged Plug in printer power cabl...

Page 19: ...int Cartridge Problem Foranyadditionalproblemsnotmentionedhere pleaserefertotheowner s manual for the printer Note The Laborie Urocap IV has a medical grade power supply available for the laptop but t...

Page 20: ...16 LABORIE Urocap IV Light Service Manual UCL SM01 V7 00 Devices not connected in app Connect add the devices to the system See owner s manual for more information...

Page 21: ...ed before the battery failure can be printed by tapping the Reprint Last Test icon Uroflowmeter signal shows vibration and or spike patterns Plastic beaker is touching the flow funnel Reposition the U...

Page 22: ...ononthebottom of the Urocap IVdevice 2 Plug in the Urocap IV charger if necessary and tap the Urocap IV icon on the Urocap IV Light UDS Light soft ware screen The icon will change to disconnected mode...

Page 23: ...Messages SYMPTOM S POSSIBLE CAUSE S CHECK CORRECTIVE ACTION S Low Battery Shutdown Is there insufficient battery power 1 Plug in the Urocap IV charger and tap the Urocap IV icon in the software The ic...

Page 24: ...mitation of Voltage Changes Voltage Fluctuations and Flicker in Public Low Voltage Supply Systems for equipment with rated current 16A per phase and not subject to conditional connection IEC61000 4 2...

Page 25: ...IV Lightsystem Increase the separation between Urocap IV Light system and the affected equipment Connect the non medical system equipment into an outlet on a circuit different from that to which the...

Page 26: ...symbol and training in ESD precautionaryprocedures The minimum contents of an ESD precautionary procedure training should include An introduction to the physics of electrostatic charge The voltage le...

Page 27: ...onic equipment RF Emissions CISPR 11 Class A The MEEQUIPMENT is suitablefor use in all establishments other than domestic and may be used in domestic establishments and those directly connected to the...

Page 28: ...nvironment Surge IEC 61000 4 5 1 kV line s to line s 2 kV line s to earth 1 kV line s to line s 2 kV line s to earth Mains power quality should be that of a typical commercial or hospital environment...

Page 29: ...each frequency range b Interference may occur in the vicinity of known RF transmitting devices and equipment marked with the following symbol IEC 61000 4 6 150 kHz to 80 MHz Radiated RF 3 V m 3 V m IE...

Page 30: ...he user of the ME EQUIPMENT can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and the ME EQUIPMENT as...

Page 31: ...This product is certified to U S and Canadian safety standards by T V S D CE Mark European Compliance Symbol Authorized Representative in the European Community Read Operator s Manual Serial Number No...

Page 32: ...CL SM01 V7 00 END USER SOFTWARE LICENSEAGREEMENT Findlicensesandterms forLABORIE productsandservicesinthefollowing locations End User License Agreement http www laborie com eula Terms and Conditions h...

Page 33: ...LABORIE Urocap IV Light Service Manual UCL SM01 V7 00...

Page 34: ...LABORIE Urocap IV Light Service Manual UCL SM01 V7 00...