LABORIE Urocap IV Light Service Manual UCL-SM01 V7.00
•
The Uroflow beaker should be rinsed and dried after use. Soap and water,
or a mild detergent solution, can be used for cleaning.
Although the beaker is reusable, it will discolor over time and may need
to be replaced.
•
The chair, Flowstand, and funnel can be wiped clean with a cloth
dampened with soap or mild detergent and wiped dry immediately.
When cleaning the funnel the maximum cleaning
temperature should not exceed 80°C (176°F).
•
Performing regular maintenance will reduce the need for costly repairs.
Check the transducer calibration of the Urocap IV every 6 months or
whenever you suspect the transducer is off calibration.
•
Pay close attention to the LED lights on the device. If they indicate a
broken connection and/or low battery, then make sure that the
connection is re-established or the battery is fully charged. For
information on LED lights, refer to the owner’s
manual.
•
Refer to
manufacturer’s
instructions for cleaning and maintenance of the
mobile device.
BATTERY: CHARGING AND PREVENTIVE MAINTENANCE
C
HARGING
T
HE
B
ATTERY
•
Lift the tab on the bottom of the Urocap IV to reveal the charging port (A).
•
Plug one end of the charging power supply cord into the charger port on the
bottom of the device and the other end into an electrical outlet.