LABORIE Urocap IV Light Service Manual UCL-SM01 V7.00
Table 7: IEC 60601-1-2:2007 Table 1 Requirements
The
ME
EQUIPMENT
is intended for use in the electromagnetic environment specified
below. The customer or the user of the
ME
EQUIPMENT
should assure that it is used in such
an environment.
Emissions Test
Compliance
Electromagnetic environment
–
guidance
RF Emissions CISPR
11
Group 1
The
ME EQUIPMENT
uses RF energy only for
its internal function. Therefore, its RF
emissions are very low and are not likely
to cause any interference in nearby
electronic equipment.
RF Emissions CISPR
11
Class A
The
ME EQUIPMENT
is suitable for use in all
establishments other than domestic, and
may be used in domestic establishments
and those directly connected to the
public low- voltage power supply
network that supplies buildings used for
domestic purposes, provided the
following warning is heeded:
Warning: This equipment/system is
intended for use by health care
professionals only. This equipment/
system may cause radio interference
or may disrupt the operation of
nearby equipment. It may be
necessary to take mitigation
measures, such as re- orienting or
relocating the
ME EQUIPMENT
or
shielding the location.
Harmonic Emissions IEC
61000-3-2
Class A
Voltage Fluctuations/
Flicker Emissions
IEC 61000-3-3
Complies