25
LABORIE Urocap IV Light Service Manual UCL-SM01 V7.00
Table 10: IEC 60601-1-2:2007 Table 6 Requirements
Recommended separation distances between portable and mobile RF communications equipment and the
The
ME EQUIPMENT
is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the
ME EQUIPMENT
can help prevent
electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the
ME EQUIPMENT
as recommended
below, according to the maximum output power of the communications equipment.
Rated maximum
output power
of transmitter
W
Separation distance according to frequency of transmitter M
150 kHz to 80 MHz
𝑑 = [
3, 5
𝑉1
] √𝑃
80 MHz to 800 MHz
𝑑 = [
3, 5
𝐸1
] √𝑃
800 MHz to 2.5 GHz
𝑑 = [
7
𝐸1
] √𝑃
0.01
0.12
0.12
0.24
0.1
0.37
0.37
0.74
1
1.17
1.17
2.34
10
3.69
3.69
7.38