Display
Operating Procedure
Description
Programming Mode
“Technical” Programming
(only on ABR espresso machine models)
Enter Password
******
Scale
Configuration
• This parameter enables the technician
to set some scale parameters like:
-
Enabling/disabling
the
offset
parameter;
- Setting the weight reading time;
- Calibrating the scales;
- Updating the scale software.
Program Scales
G1 Scale
PRESENT
5
This parameter indicates that the scale is properly connected; navigate using the
T1
and T2
buttons
to display the next menu.
Calibrate
G1 Scale
6
Press the T3 button
to start the calibration procedure.
Empty G1 Scale
And Press Enter
Place 100g on G1
And Press Enter
7
Remove any object from the scale, then press the T3 button
to confirm.
Place the reference weights onto the scale, then press the T3 button
to confirm. At the end of the process,
the self-calibration values or a confirmation message are displayed.
Repeat this operation for each group.
G1 Scale: v0.0.0
Upgrade to 0.0.0
8
Press
T2
to proceed with setting and press the T3 button
to update the scale firmware if necessary.