KEW6305
11.2
11.2 Auto-ranging
Auto-ranging function is available at W, Wh, DEMAND and WAVE Ranges. Current values in wide
range can be measured with this function; it is helpful when load capacitances vary dramatically
according to day and time.
●
Range: 2-range-auto/ max and min range of each Clamp sensor
●
Range shift to upper one when crest values equal to twice as much as F.S (sine wave) at min
range is detected.
Accurate values may not be obtained when substantial fluctuations in 1 sec.
11.3 Operation at AC power interruption
When an AC power supply is interrupted during recording, KEW6310 operates as follows.
●
Power supply : restores to battery when batteries have been installed
●
Measurement data : saved until the last interval before an interruption
●
Operation after interruption : recording restarts with preset settings if a power is
interrupted during recording. In this case, occurrence of interruption is recorded with
time and date information. (STOP) Restoration is also recorded as well. (START)
Instrument doesn’t power on again automatically when a power interruption occurs
and restores other than recording period.
Files in CF card or Internal memory may be destroyed if an AC power is interrupted while
accessing to them.
Use of AC power supply and batteries at the same time is recommended if power interruptions
are concerned.
11.4 Data check
Past 10 data (including the latest one) can be recalled and checked on the LCD.
Select the
DATA CHECK
Range for checking the data.
Data No.
01
02
・・・
09 10
Saved data
Latest
data
Two before
the latest
Nine before
the latest
Ten before
the latest