3 Operation Method - Full-Auto Protocol
36
QuickGene-Auto240L Operation Manual
3
Note:
Be sure to set a waste container in empty status.
Operation with residual waste fluid in the waste fluid container may cause an overflow of
waste fluid. Dispose of the fluid before setting a container in the holder.
(3)
Set a reagent container holder in a reagent container holder slot.
(4)
Set awash buffer bottle in the wash buffer bottle rack in the drawer (at setting position No.4).
Note:
If the setting is not complete, the inability to absorb wash buffer may influence the operation
result.
Abide by the following and set a wash buffer bottle correctly:
・
The wash buffer bottle is set in such a way that the opening comes to the left side of the rack.
・
The ends of 2 inlet tubes must reach the bottom of the wash buffer bottle.
・
The inlet tubes should not be kinked midway.
4
Inlet Tube
Wash Buffer
Bottle
Wash Buffer
Bottle Rack