2
Basic Functions
16
QuickGene-Auto240L Operation Manual
2
2.3
Safety Mechanisms
The safety mechanisms of this system are explained below.
This system is equipped with the following safety mechanisms.
・
Sliding door, flap door
・
Earth leakage breaker
・
Agitation cover
・
Operation panel, buzzer
■
Sliding Door, Flap Door
After power activation, the sliding doors and flap doors are locked. Temporary unlocking or door
opening/closing is possible after power activation via panel operation, but panel operation is restricted while in
unlocked status.
The opening and closing of doors are monitored with a sensor, and operations such as the isolation process are
prohibited while a door is open.
■
Earth Leakage Breaker
When an overcurrent occurs due to earth leakage or system malfunction, the earth leakage breaker will shut off
power.
The breaker condition can be checked on the breaker switch by opening the drawer of the system.
“ON” is seen while the breaker switch is in normal status, and “OFF” is seen when the circuit is shut off.
If the earth leakage breaker is shut off, please contact our customer consultation desk.
■
Agitator Cover
The system lysate unit is equipped with an open/close type cover to prevent the ejection of the lysate tube during
automatic operation.
■
Operation Panel, Alarm
The system operating condition is indicated using the operation panel and the alarm.
Summary of Contents for QuickGene-Auto240L
Page 1: ...Automated Nucleic Acid Isolation System QuickGene Auto240L Ver 1 0 1 Operation Manual ...
Page 2: ......
Page 14: ...xii QuickGene Auto240L Operation Manual ...
Page 44: ...2 Basic Functions 30 QuickGene Auto240L Operation Manual 2 ...
Page 78: ...3 Operation Method Full Auto Protocol 64 QuickGene Auto240L Operation Manual 3 ...