![Kurabo QuickGene-Auto240L Operation Manual Download Page 15](http://html.mh-extra.com/html/kurabo/quickgene-auto240l/quickgene-auto240l_operation-manual_2011195015.webp)
1 System Installation Procedures
1
QuickGene-Auto240L Operation Manual
1
1 System Installation Procedures
The installation procedures for this system are explained below.
1.1 Conditions for Installation, Operating Environment
The conditions for system installation and the operating environment are explained here.
■ Conditions for Installation
・
Do not install this system where water may be poured or spilt, which may
failure, fire or electric shock.
・
Be sure to remove the power plug from the outlet whenever this system is
moved. Watch your step carefully while operating. Injury, fire or electric
may result from a damaged power cable.
・
Do not install this system in a location with intense vibration or at an unstable
incline. This may cause injury or failure.
・
Do not install this system in a location exposed to direct sunlight or in the
proximity of a heating or cooling appliance. This may cause a shortened
lifetime or failure.
Please install this system in an indoor location as shown below for normal, safe use.
・
A location with the following secure space for installation.
Required footprint: Width
≧
178 cm × Depth
≧
152 cm × Height
≧
205 cm (when placed on an
80-cm high experiment table)
Warning
:
Caution:
10cm
(40 cm)
(20 cm)
(10 cm)
(70 cm)
(10 cm)
205 cm
178 cm
152 cm
Summary of Contents for QuickGene-Auto240L
Page 1: ...Automated Nucleic Acid Isolation System QuickGene Auto240L Ver 1 0 1 Operation Manual ...
Page 2: ......
Page 14: ...xii QuickGene Auto240L Operation Manual ...
Page 44: ...2 Basic Functions 30 QuickGene Auto240L Operation Manual 2 ...
Page 78: ...3 Operation Method Full Auto Protocol 64 QuickGene Auto240L Operation Manual 3 ...