OPERATION
6
45
H250 M40
www.krohne.com
03/2020 - 4000640706 - MA H250-M40 R06 en
6.5 Error messages ESK4-T
Error messages and warnings are indicated by one of the following symbols in the bottom left
corner of the display. The
^
^
^
^
key switches from the measured value display to the display of
current errors / warnings. The table below contains a description of possible error messages.
Symbol
NE107
Category
Description
Consequence
F
F
F
F
Failure
No measured value available
Output signal invalid.
Error current signal is output.
S
Out of specification
Measurement available but increased measurement
uncertainty. Device must be checked.
M
Maintenance required
Measurement still sufficiently accurate but
maintenance required.
C
Function check
Device is in test or calibration mode. The output signal
temporarily does not correspond to the measurement.
I
Information
No influence on the measurement, information only.
Table 6-8: Description of the symbols
Error message
Description
Category
Remedy
NOT LINEARIZED
Linearization faulty or not
activated = measuring
error.
S
Activate linearization or carry it out again (HART
®
communication and linearization software are
required; the original calibration values must be
known), or send the device back to the
manufacturer for linearization.
NEW LINEARI. TABLE
BAD
Faulty or missing data in
the linearization table =
measuring error.
S
1
LINEARIZATIO UNDER
CONFIG
The device is in
linearization mode =
measuring error.
S
Complete the linearization and activate it (HART
®
communication and linearization software are
required), or send the device back to the
manufacturer for linearization.
UNIT SYSTEM
CONFLICT
The unit for the
linearization flow is
incompatible with the
selected flow type
(mass/volume).
S
Correct error, carry out linearization again if
necessary (HART
®
communication and
linearization software are required), or send the
device back to the manufacturer for linearization.
TOO FEW ENTRIES
The linearization table has
too few data points.
S
Carry out linearization at at least 5 points (HART
®
communication and linearization software are
required), or send the device back to the
manufacturer for linearization.
NOT MONOTONOUS
The sequence of the
linearization values is not
strictly monotonic
increasing.
S
Check linearization and/or carry it out again
(HART
®
communication and linearization
software are required), or send the device back to
the manufacturer for linearization.
FIRST NOT 0 %
The first flow value if the
linearization table is not
0%.
LAST NOT 100 %
The last flow value if the
linearization table is not
100%.