4
ELECTRICAL CONNECTIONS
38
H250 M40
www.krohne.com
03/2020 - 4000640706 - MA H250-M40 R06 en
4.3 Grounding connections
4.4 Ingress protection
The measuring device meets all requirements of ingress protection IP66/68.
Therefore it is essential to observe the following points.
•
Use only original gaskets. They must be clean and free of any damage. Defective gaskets must
be replaced.
•
The electrical cables must be undamaged and must comply with regulations.
•
The cables must be laid with a loop
3
upstream of the measuring device to prevent water
from getting into the housing.
•
The cable feedthroughs
2
must be tightened.
•
Close the unused cable feedthroughs using blanking plugs
1
.
Figure 4-15: Grounding connections
1
Grounding connection on the indicator
2
Outer grounding connection
DANGER!
The grounding wire may not transfer any interference voltage.
Do not use this grounding cable to ground any other electrical devices.
DANGER!
After all servicing and maintenance work on the measuring device, the specified protection class
must be ensured again.
Figure 4-16: Laying the cable
1
Use blanking plugs if no cable is routed through
2
Tighten cable feedthrough firmly
3
Lay the cable in a loop