Sonar room requirements
Topics
Environmental requirements, page 17
Size and access requirements, page 17
Requirements for insulation, heating and ventilation, page 18
Requirements for electrical installations, cables and communication, page 18
Environmental requirements
The EM 712 units must be installed in a dry and dust-free environment. The units are not
fully protected against humidity, dust or water.
It is important that the sonar room is kept dry. The EM 712 units must not be exposed
to excessive temperatures, moisture or humidity. Such conditions can cause corrosive
attacks and subsequent failures to the electronic circuitry. Visit the sonar room at
regular intervals to check temperature and humidity, and take the necessary actions
if the environmental conditions are poor.
Observe the environmental specifications related to the EM 712 units.
Related topics
Environmental requirements, page 161
Size and access requirements
A well designed sonar room with a well fitted size and easy access reduces the risk of
corrosion, and simplifies maintenance. This increases system reliability.
The sonar room must be large enough to house all the system units. The room must
provide enough space to allow efficient maintenance. You must be able to keep all the
cabinet doors fully open without undue restriction to your movements.
1
The room must not be used for any other heavy machinery.
2
The room must not be unnecessarily obstructed by girders, pipes etc, which may
cause installation problems or impede maintenance.
3
The sonar room must be accessible under all conditions at sea or at a berth.
4
All doors or hatches must be designed so that the tools and equipment can be
removed without being disassembled.
Related topics
Summary of Contents for EM 712
Page 71: ...401027 B 71 216148 EM 712 Transducer TX1 dimensions Drawing file ...
Page 72: ...72 401027 B Kongsberg EM 712 Installation manual ...
Page 73: ...401027 B 73 221048 EM 712 Transducer TX2 dimensions Drawing file ...
Page 74: ...74 401027 B Kongsberg EM 712 Installation manual ...
Page 75: ...401027 B 75 219621 EM 712 Transducer RX1 dimensions Drawing file ...
Page 76: ...76 401027 B Kongsberg EM 712 Installation manual ...
Page 77: ...401027 B 77 216146 EM 712 Transducer RX2 dimensions Drawing file ...
Page 78: ...78 401027 B Kongsberg EM 712 Installation manual ...
Page 79: ...401027 B 79 223137 EM 712 Transducer mounting frame 0 5 degrees Drawing file ...
Page 80: ...80 401027 B Kongsberg EM 712 Installation manual ...
Page 81: ...401027 B 81 223139 EM 712 Transducer mounting frame 1 degree Drawing file ...
Page 82: ...82 401027 B Kongsberg EM 712 Installation manual ...
Page 83: ...401027 B 83 223273 EM 712 Transducer mounting frame 2 degrees Drawing file ...
Page 84: ...84 401027 B Kongsberg EM 712 Installation manual ...
Page 85: ...401027 B 85 317812 EM 712 Casing w mounting frame 0 5 degrees Drawing file ...
Page 87: ...401027 B 87 375817 EM 712 Combined casing w mounting frame 1 degree Drawing file ...
Page 89: ...401027 B 89 396402 EM 712 Transmitter Unit dimensions Drawing file ...
Page 91: ...401027 B 91 396428 EM 712 Receiver Unit dimensions Drawing file ...
Page 92: ...92 401027 B 385422 Processing Unit dimensions Kongsberg EM 712 Installation manual ...
Page 93: ...401027 B 93 378828 Hydrographic Work Station dimensions Drawing file ...
Page 94: ...94 401027 B Kongsberg EM 712 Installation manual ...
Page 95: ...401027 B 95 371591 Rack installation kit dimenisons Drawing file ...
Page 96: ...96 401027 B 370275 Remote Control Unit K REM dimensions Kongsberg EM 712 Installation manual ...
Page 97: ...401027 B 97 Drawing file ...
Page 163: ...401027 B 163 Certificates IEC 60945 IACS E10 Technical specifications ...
Page 177: ...401027 B 177 7 Observe the handling rules for transducers Equipment handling ...
Page 188: ... 2018 Kongsberg Maritime ...