Key Operator Instructions
7E6174 – 16FEB01
5 –51
2. Insert the end of the Developer, or Fixer Replenishment Hose into a Graduated
Cylinder.
3. Hold the Graduated Cylinder over the Wash Section of the Processor to prevent
spills from contaminating the processing solutions.
4. Press and hold the Replenishment Calibration Switch for 5 seconds. The switch
is located on the side wall of the Processor Tank.
5. Allow the Replenishment Pumps to turn on and deliver the preset volume of
replenishment solution.
6. Compare the volume of replenishment solution in the Graduated Cylinder of the
replenishment volume you set in the previous procedure.
7. Install the Wet Section Cover and carefully push in the Processor.
Warning
See Step 9 on 4 –11.
If the volumes are not within 10% of each other, do the "Calibrating the
Replenishment System” procedure described on Page 5– 70.
H186_1087AC
H186_1087ACA
Fixer
Replenishment
Hose
Developer
Replenishment
Hose
Graduated
Cylinder
H186_1087AC
Switch
Replenishment
Calibration
H186_1087ACB