Section 2 -- Installation
2-49
2005 July Rev. B 7F3318
2-11-3. Selecting Parameters for MODALITIES that are not DICOM-GSDF-Compliant
1. Refer to Table 2-4 and find the type of MODALITY for which the IMAGER is being configured.
2. Select the TFT Set recommended in Table 2-4 for this type of MODALITY. If more than one set is
listed, read the notes in the table for guidance in selecting the appropriate set.
3. Select the Table No. (contrast setting) recommended for the chosen TFT Set. If a number is in “bold,”
select this number first (You may have to try more than one Table No, during the configuration
process.)
4. Select the Image Density associated with the Table No. you chose. If there is a range of densities:
•
Select the density value indicated in bold, if there is one.
•
If there is no bold value, select the middle value. For example, if the range is 2.7 to 3.1, select 2.9.
•
If there are only 2 density values, select either one.
5. Select the Smoothing Type recommended in Table 2-4.
6. Enter the selected parameter values on the “Image Optimization” screen in the SERVICE TOOL.
2-11-4. Selecting Parameters for DICOM GSDF-Compliant MODALITIES
Note
This procedure does not apply to GSDF-Compliant MODALITIES that use PLUTs for
all
images.
1. Use TFT Set GSDF0.87.
2. Refer to Table 2-4 and find the type of MODALITY for which the IMAGER is being configured.
3. Select the Density (Dmax) recommended for this type of MODALITY. If there is a range of densities:
•
Select the density value indicated in bold, if there is one.
•
If there is no bold value, select the middle value. For example, if the range is 2.7 to 3.1, select 2.9.
•
If there are only 2 density values, select either one.
4. Refer to Table 2-5 and select the Table No. (contrast setting) that corresponds to the chosen Density
(Dmax) value.
5. Refer to Table 2-4 and select the Smoothing Type for the selected type of MODALITY.
6. Enter the selected parameter values on the “Image Optimization” screen in the SERVICE TOOL.
Summary of Contents for DryView 8150
Page 1: ...SERVICE MANUAL Kodak DryView 8150 LASER IMAGER 1961259 7F3318 07 05 ...
Page 20: ...Service Manual 1 6 7F3318 2005 July Rev B BLANK PAGE ...
Page 34: ...Service Manual 2 14 7F3318 2005 July Rev B 6 Remove the 2 SCREWS and 2 TAGS 6 ...
Page 50: ...Service Manual 2 30 7F3318 2005 July Rev B 6 Check each parameter ...
Page 91: ...Section 2 Installation 2 71 2005 July Rev B 7F3318 BLANK PAGE ...
Page 105: ...Service Manual 3 14 7F3318 2005 July Rev B BLANK PAGE ...
Page 182: ...Section 4 Disassembly Reassembly 4 77 2005 July Rev B 7F3318 BLANK PAGE ...
Page 196: ...Service Manual 5 14 7F3318 2005 July Rev B Figure 5 9 Cleaning the HEAT SHIELD and ROLLERS ...
Page 200: ...Service Manual 5 18 7F3318 2005 July Rev B Figure 5 11 Adjusting the DIVERTER ...
Page 236: ...Service Manual 6 34 7F3318 2005 July Rev B Figure 6 17 Pickup Operation Sheet 1 of 3 ...
Page 238: ...Service Manual 6 36 7F3318 2005 July Rev B Figure 6 17 Pickup Operation Sheet 3 of 3 ...
Page 240: ...Service Manual 6 38 7F3318 2005 July Rev B Figure 6 18 Load Platen Operation Sheet 2 of 2 ...
Page 241: ...Section 6 Theory of Operation 6 39 2005 July Rev B 7F3318 Figure 6 19 Expose Operation ...
Page 242: ...Service Manual 6 40 7F3318 2005 July Rev B Figure 6 20 Unload Platen Operation ...
Page 244: ...Service Manual 6 42 7F3318 2005 July Rev B Figure 6 21 Processor Operation Sheet 2 of 2 ...
Page 246: ...Service Manual 6 44 7F3318 2005 July Rev B Figure 6 22 Front Door Open Operation Sheet 2 of 2 ...
Page 252: ...Service Manual 6 50 7F3318 2005 July Rev B Figure 6 27 Sequence of AIQC and GSM Functions ...
Page 444: ...Service Manual 8 102 7F3318 2005 July Rev B BLANK PAGE ...
Page 466: ...Service Manual 9 22 7F3318 2005 July Rev B BLANK PAGE ...
Page 468: ...Service Manual 10 2 7F3318 2005 July Rev B BLANK PAGE ...
Page 469: ......