A-61560 January 2007
103
Cleaning or replacing the
separation module
The separation module should be removed from the scanner for best
cleaning results.
NOTE: When removing or cleaning the separation module, it may be
easier to access the separation module if you remove the input
tray. To remove the input tray, gently push the input tray to the
left or right to release the pin from the slot, and lift it out of
position.
1. Push the separation module release lever down and lift the
separation module up and out of position.
2. Manually rotate and wipe the separation module tires with a roller
cleaning pad.
3. Inspect the separation module.
If the separation module tires show signs of wear or damage,
replace the tires or the separation module. See the next section for
procedures.