![Knick Stratos Pro A2 COND Series User Manual Download Page 79](http://html.mh-extra.com/html/knick/stratos-pro-a2-cond-series/stratos-pro-a2-cond-series_user-manual_1980189079.webp)
79
Step
Action/Display
Remark
Select HOLD
mode
Press menu key to call
the selection menu,
select HOLD using the
keys, press enter
to confirm.
Now the device is in HOLD
mode. The HOLD mode can also
be activated externally via the
HOLD input. During HOLD the
output current is frozen at its last
value or set to a fixed value.
Disconnect old
sensor,
connect new
sensor.
Temporary messages are dis-
played during the replacement
but neither output to the alarm
contact nor entered in the log-
book.
Wait until the
sensor data are
displayed.
MEMO SENS
Check sensor
data
MEMO SENS
View sensor informa-
tion using
keys,
press enter to confirm.
You can view the sensor manu-
facturer and type, serial number,
and last calibration date.
Check
measured
values,
then
exit HOLD.
Hit meas key:
Return to the selec-
tion menu. Hold meas
key depressed: Device
switches to measuring
mode.
The sensor replacement is
entered in the extended
logbook.
Digital Sensors
Summary of Contents for Stratos Pro A2 COND Series
Page 121: ...121 Buffer Tables 10 DIN 19267 buffers ...
Page 141: ...141 Index ...