Configuration
Menu item
Action
Choices
Temperature
compensation,
process medium
For pH measurement
only: Select temperature
compensation of the
process medium.
Linear: LIN
Table: USERTAB
Select using
key,
confirm by pressing
enter.
Temperature
compensation, linear
Enter the linear tempera-
ture compensation of the
process medium.
Enter value using
keys.
Press enter to confirm.
Temperature
compensation, table
Confirm safety
prompt,
then enter values
(5°C step size)
When you have selected
temperature compensa-
tion via table (USERTAB),
you can enter values for
a TC table from 0 to 95 °C
in 5-K steps.
The analyzer displays
temperature values in
5 °C steps. You must spec-
ify the percent deviation
of the measured value
from each of these tem-
perature values.
Intermediate values are
linearly interpolated.
TC compensation can be
defined separately for
parameter sets A and B.
0 ... 95 °C
(5°C step size)
5