Step
Action/Display
Remark
Select buffer set -U1-
(CONFIG / SNS menu)
Select buffer solution
1 for editing
Select “YES“ using up/
down key.
You are prompted for
confirmation to prevent
accidental changes of
the settings.
Editing the values of
buffer solution 1
Edit using arrow keys,
press enter to confirm and
proceed to next tempera-
ture value.
Enter the values for the
first buffer solution in
5°C steps.
The difference to the
next value must not
exceed 0.25 pH unit.
Select buffer solution
2 for editing
The difference between
buffer solutions for iden-
tical temperatures must
be greater than 2 pH
units.
-U1- Specifiable Buffer Set