Service
Service (SERVICE) A
In the Service mode you can access the following menus:
MONITOR
Displaying currently measured values
SENSOR
Resetting TTM (ISM only ),
incrementing the autoclaving counter
OUT1
Testing current output 1
OUT2
Testing current output 2
CODES
Assigning and editing passcodes
DEFAULT
Resetting the device to factory settings
OPTION
Enabling options via TAN.
Note:
HOLD is active during Service mode!
Action
Key/Display
Remark
Activate
Service
Press menu key to call the selection
menu.
Select SERVICE using
keys,
press enter to confirm.
Passcode
Enter passcode “5555“ for service
mode using the
keys.
Press enter to confirm.
Display
In service mode the following icons
are displayed:
• [diag] mode indicator
• HOLD triangle
• Service (wrench)
Exit
meas
Exit by pressing meas.